BRILLIANCE 40 SLICE CT SCANNER
Report
- Report Number
- 1525965-2007-00045
- Event Type
- Malfunction
- Date Received
- December 21, 2007
- Date of Event
- December 12, 2007
- Report Date
- December 13, 2007
- Manufacturer
- PHILIPS MEDICAL SYSTEMS
- Product Code
- JAK
- PMA / PMN Number
- K033326
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
ON MAY 25, 2007, PHILIPS MEDICAL SYSTEMS TOOK ACTION TO ADVISE THE INSTALLED BASE OF THIS MATTER THROUGH A PRODUCT SAFETY NOTIFICATION AND PHILIPS MEDICAL SYSTEMS WILL UPDATE THE CUSTOMER'S SYSTEMS TO MITIGATE THIS ISSUE. THIS EVENT IS SIMILAR TO MDR REPORT 1525965-2007-00006 AND A DECISION WAS REACHED TO FILE AN MDR ON THIS EVENT. THIS ISSUE IS ASSOCIATED WITH CUSTOMER COMPLAINT.
IT WAS REPORTED THAT AN ARTIFACT VISUALIZED AS A SMALL PATTERN OF FOUR LOW CONTRAST DOTS IS PRESENT IN THE BRAIN SCAN. THE PATIENT WAS MISDIAGNOSED WITH A BRAIN BLEED BECAUSE OF THIS ARTIFACT. AFTER FURTHER EVALUATION, THE IMAGE WAS DETERMINED TO BE NON DIAGNOSTIC. NO ADVERSE CONSEQUENCES WERE REPORTED. ORIGINALLY THE NIGHT READING SERVICE READ PATIENT TO HAVE A BLEED. PATIENT WAS PUT UNDER OBSERVATION AND LATER DIAGNOSIS CORRECTION WAS MADE TO NO BLEED. NO SURGICAL TEAM READINESS OR PREPARATION WAS MADE. RADIOLOGISTS FEEL THE ARTIFACT COULD BE MISTAKEN FOR PATHOLOGY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BRILLIANCE 40 SLICE CT SCANNER | COMPUTED TOMOGRAPHY SCANNER | JAK | PHILIPS MEDICAL SYSTEMS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 1 MO |