T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY
Report
- Report Number
- 3013756811-2020-13010
- Event Type
- Malfunction
- Date Received
- February 10, 2020
- Date of Event
- December 1, 2019
- Report Date
- February 10, 2020
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- OZO
- UDI-DI
- 00853052007264
- PMA / PMN Number
- P180008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- OTHER
Narratives
PER TANDEM¿S T:SLIM X2 G6 USER GUIDE: ¿DO NOT USE ANY OTHER INSULIN WITH YOUR SYSTEM OTHER THAN U-100 HUMALOG OR U-100 NOVOLOG. ONLY HUMALOG AND NOVOLOG HAVE BEEN TESTED AND FOUND TO BE COMPATIBLE FOR USE IN THE SYSTEM." NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT MULTIPLE INTERMITTENT OCCLUSION ALARMS OCCURRED. CUSTOMER'S BLOOD GLUCOSE RANGED FROM 100-413 MG/DL. PRIOR TO TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT, THE CUSTOMER REVERTED TO AN ALTERNATE PUMP FOR INSULIN THERAPY. WHEN THE CUSTOMER CONNECTED BACK TO THE TANDEM INSULIN PUMP, OCCLUSION ALARMS CONTINUED. REPORTEDLY, THE PUMP SUPPLIES WERE CHANGED TO ADDRESS THE ISSUE. ADDITIONALLY, CUSTOMER WAS USING APIDRA INSULIN. TANDEM TECHNICAL SUPPORT EDUCATED CUSTOMER REGARDING CARTRIDGE/INSULIN LABELING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 149147 | T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY | AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND | OZO | TANDEM DIABETES CARE | 1000096 | 00853052007264 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 19 YR |