840 VENTILATOR
Report
- Report Number
- 8020893-2020-00020
- Event Type
- Malfunction
- Date Received
- February 10, 2020
- Report Date
- February 10, 2020
- Manufacturer
- COVIDIEN
- Product Code
- CBK
- PMA / PMN Number
- K151252
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER
Narratives
DEVICE EVALUATION SUMMARY: THE MEDTRONIC FIELD SERVICE ENGINEER EVALUATED THE VENTILATOR AND REPLACED THE BREATH DELIVERY (BD) PRINTED CIRCUIT BOARD (PCB), PIEZO ALARM, ANALOG INTERFACE (AI) PCB AND THE POWER SUPPLY. THE VENTILATOR PASSED ALL TESTING PER MANUFACTURER SPECIFICATIONS AT THE TIME OF SERVICE AND WAS RETURNED TO THE CUSTOMER. THE FOLLOWING COMPONENTS WERE RETURNED FOR FURTHER ANALYSIS: AI PCB, BD PCB, CONDOR POWER SUPPLY THE AI PCB WAS VISUALLY INSPECTED AND NOTED MULTIPLE COMPONENTS WERE THERMALLY DAMAGED INCLUDING DIODES D58, D59, D60, D62 AND D63. THE PCB WAS NOT INSTALLED IN THE VENTILATOR TO AVOID FURTHER DAMAGE. THE COMPONENT DAMAGE WAS CAUSED BY THE CONDOR POWER SUPPLY WHICH WAS ALSO RETURNED UNDER THIS COMPLAINT NUMBER. THE BD PCB WAS VISUALLY INSPECTED AND THE FOLLOWING WAS OBSERVED: CAPACITOR C100 WAS PHYSICALLY BROKEN OFF THE PCB. THIS WAS NOT THE CAUSE OF THE REPORTED FAULT. THE DAMAGE OCCURRED DUE TO MISHANDLING AT SOME TIME DURING THE 20 YEAR SERVICE LIFE OF THIS PCB. THE REPORTED FAULT WAS CAUSED BY THE CONDOR POWER SUPPLY WHICH WAS ALSO RETURNED UNDER THIS COMPLAINT NUMBER. THE CONDOR POWER SUPPLY WAS RETURNED AND THE FOLLOWING WAS FOUND: THERE WAS AN ODOR OF THERMAL COMPONENT DAMAGE FROM THE UNIT. AN INTERNAL VISUAL INSPECTION REVEALED THAT THE C30 CAPACITOR DEVICE WAS SEVERELY THERMALLY DAMAGED. THE C30 DECOUPLING CAPACITOR IS A 4700 PF, 1KV CERAMIC DEVICE,CONNECTED BETWEEN THE T4 TRANSFORMER AND CHASSIS GROUND. THE DEVICE INITIALLY BROKE DOWN INTERNALLY CAUSING A SHORT CIRCUIT TO CHASSIS GROUND BEFORE IT BURNED UP. THIS CAUSED FURTHER DAMAGE TO OTHER PCBS ON THE VENTILATOR. THE POWER SUPPLY MANUFACTURE DATE: AUGUST 1998. THIS UNIT HAS A SERVICE LIFE OF APPROXIMATELY 20 YEARS. IF THIS FAILURE OCCURRED DURING VENTILATION THE VENTILATOR WOULD GENERATE A VENT INOP CONDITION. THE APPROPRIATE AUDIO AND VISUAL ALARMS WOULD ENUNCIATE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT AN 840 VENTILATOR WOULDN'T TURN ON. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT. BASED ON PRODUCT ANALYSIS OF THE POWER SUPPLY, THE POWER SUPPLY WOULD HAVE CAUSED A LOSS OF VENTILATION IF IT WERE TO OCCUR ON A PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 148657 | 840 VENTILATOR | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | COVIDIEN | 4-076028-00 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |