FDA Adverse Event Malfunction Summary report: N

FISCHER CONE BIOP EX MED

MDR report key: 9687490 · Received February 10, 2020

Report

Report Number
1216677-2020-00002
Event Type
Malfunction
Date Received
February 10, 2020
Date of Event
November 12, 2019
Report Date
July 27, 2020
Manufacturer
COOPERSURGICAL, INC.
Product Code
HGI
PMA / PMN Number
K061651
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

*INVESTIGATION. X-INITIATED MANUFACTURER'S INVESTIGATION. X-NO SAMPLE RETURNED. X-REVIEW DHR. *ANALYSIS AND FINDINGS. REF (B)(4). DISTRIBUTION HISTORY THE 15006-02 FISHER CONE ASSEMBLY WAS PURCHASED FROM GEOTEC, INC. AS AN OEM FINISHED PRODUCT, ISSUED TO WORK ORDER 219147 AS CSI PART NUMBER 900-151 AND COMPLETED 5/7/2018. MANUFACTURING RECORD REVIEW DHR-900-151 - 238680 WAS REVIEWED AND NO NON-CONFORMITIES, RELATED TO THE COMPLAINT CONDITION, WERE NOTED. INCOMING INSPECTION REVIEW IQC- 18-05-01-004 WAS REVIEWED AND NO NON-CONFORMITIES, RELATED TO THE COMPLAINT CONDITION, WERE NOTED. SERVICE HISTORY RECORD. SERVICE HISTORY NOT APPLICABLE FOR THIS PRODUCT. HISTORICAL COMPLAINT REVIEW. A REVIEW OF THE 2-YEAR COMPLAINT HISTORY DID SHOW SIMILAR REPORTED COMPLAINT CONDITIONS. PRODUCT RECEIPT. THE COMPLAINT PRODUCT HAS NOT BEEN RETURNED TO COOPERSURGICAL. VISUAL EVALUATION. EVALUATION OF THE COMPLAINT PRODUCT COULD NOT BE COMPLETED AS THE COMPLAINT PRODUCT HAS NOT BEEN RETURNED TO COOPERSURGICAL. IF THE PRODUCT SHOULD BE RETURNED AT A LATER DATE, IT WILL BE EVALUATED, AND ANY FINDINGS WILL BE APPENDED TO THIS INVESTIGATION. FUNCTIONAL EVALUATION. EVALUATION OF THE COMPLAINT PRODUCT COULD NOT BE COMPLETED AS THE COMPLAINT PRODUCT HAS NOT BEEN RETURNED TO COOPERSURGICAL. IF THE PRODUCT SHOULD BE RETURNED AT A LATER DATE, IT WILL BE EVALUATED, AND ANY FINDINGS WILL BE APPENDED TO THIS INVESTIGATION. ROOT CAUSE. DEFINITIVE ROOT CAUSE IS INDETERMINABLE HOWEVER, PREVIOUS TESTING PERFORMED IN 2011 IN TRYING TO REPLICATE REPORTED EVENTS OF THE WIRE "BURNING" OR "SNAPPING" INDICATED THE PRODUCT PERFORMED AS INTENDED, THE TESTING WAS REPEATED IN MARCH OF 2019 AND RESULTED IN THE SAME MANNER. *CORRECTION AND/OR CORRECTIVE ACTION CORRECTIVE ACTION: NO FURTHER CORRECTIVE ACTION IS NECESSARY, AS THE COMPLAINT CONDITION WAS NOT CONFIRMED. *PREVENTATIVE ACTION ACTIVITY . COOPERSURGICAL WILL CONTINUE TO MONITOR THIS COMPLAINT CONDITION FOR TRENDS.

Description of Event or Problem · 0

PER MW5091773 RECEIVED - REPORT STATED " PATIENT CONSENTED FOR A LAPAROSCOPIC TUBAL LITIGATION, REMOVAL OF IUD AND LEEP OF THE CERVIX, DURING THE LEEP, THE TIPS OF THE ELECTRODE CAME OFF THE ELECTRODE TO THE PATIENT UTERUS. A HYSTERECTOMY WAS DONE TO LOCATE THE TIP OF THE ELECTRODE AND IT WAS RETRIEVED. PLEASE SEE ATTACHED MW5091773 FOR DETAILS. UPDATE -01.29.2020 - INITIATED FOLLOW-UP REVIEWED EXTRA STEP TO COMPLETE PROCEDURE WAS HYSTEROSCOPY. NOT HYSTERECTOMY AS REPORTED ON MW5091773. REF (B)(4).

Additional Manufacturer Narrative · 1

COOPERSURGICAL INC. IS CURRENTLY INVESTIGATING THE REPORTED COMPLAINT CONDITION. ONCE THE INVESTIGATION IS COMPLETED A FOLLOW-UP REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

PER MW5091773 RECEIVED - REPORT STATED " PATIENT CONSENTED FOR A LAPAROSCOPIC TUBAL LITIGATION, REMOVAL OF IUD AND LEEP OF THE CERVIX, DURING THE LEEP, THE TIPS OF THE ELECTRODE CAME OFF THE ELECTRODE TO THE PATIENT UTERUS. A HYSTERECTOMY WAS DONE TO LOCATE THE TIP OF THE ELECTRODE AND IT WAS RETRIEVED. PLEASE SEE ATTACHED MW5091773 FOR DETAILS. UPDATE -01.29.2020 - INITIATED FOLLOW-UP REVIEWED EXTRA STEP TO COMPLETE PROCEDURE WAS HYSTEROSCOPY NOT HYSTERECTOMY AS REPORTED ON MW5091773 (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
153991 FISCHER CONE BIOP EX MED FISCHER CONE BIOP EX MED HGI COOPERSURGICAL, INC. 900-151 238680

Patients

Seq Age Sex Outcome Treatment
1 Other