FDA Adverse Event Malfunction Summary report: N

DRIVE SHAFT-MINIMUM 520MM LENGTH-FOR USE WITH RIA

MDR report key: 9684696 · Received February 7, 2020

Report

Report Number
2939274-2020-00726
Event Type
Malfunction
Date Received
February 7, 2020
Report Date
January 20, 2020
Manufacturer
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
Product Code
HTO
UDI-DI
10886982189042
PMA / PMN Number
K042899
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H10 ADDITIONAL NARRATIVE: D4: LOT NUMBER UPDATED. H3, H6: INVESTIGATION SUMMARY. INVESTIGATION SUMMARY. INVESTIGATION FLOW: DAMAGE . VISUAL INSPECTION: THE DRIVE SHAFT-MINIMUM 520MM LENGTH-FOR USE WITH RIA (P/N: 314.743, LOT NUMBER: 14382-01) WAS RECEIVED AT US CQ. UPON VISUAL INSPECTION, THE DEVICE IS BROKEN AT THE DISTAL END. DEVICE FAILURE/DEFECT IDENTIFIED? YES. DIMENSIONAL INSPECTION: NO DIMENSIONAL INSPECTION WAS PERFORMED DUE TO POST-MANUFACTURING DAMAGE. DOCUMENT/SPECIFICATION REVIEW: NO DESIGN ISSUES OR DISCREPANCIES WERE IDENTIFIED. COMPLAINT CONFIRMED? YES. INVESTIGATION CONCLUSION: THIS COMPLAINT IS CONFIRMED AS THE DISTAL END OF THE DEVICE IS BROKEN. NO DEFINITIVE ROOT CAUSE COULD BE DETERMINED BASED ON THE PROVIDED INFORMATION. NO NEW, UNIQUE OR DIFFERENT PATIENT HARMS WERE IDENTIFIED AS A RESULT OF THIS EVALUATION. THERE WAS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE CONTRIBUTED TO THE COMPLAINT. NO DESIGN ISSUES WERE OBSERVED DURING THE DOCUMENT/SPECIFICATION REVIEW. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT PART # 314.743. SYNTHES LOT # 5236556. SUPPLIER LOT # 14382-01. RELEASE TO WAREHOUSE DATE: MAY 04, 2006. SUPPLIER: CRITERION TOOL & DIE, INC. NO NCR'S WERE GENERATED DURING PRODUCTION. DEVICE HISTORY REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THIS PRODUCT, AND ANY SUB-COMPONENTS, WHICH WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

ADDITIONAL DEVICE PRODUCT CODE: HRX. COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION, BUT HAS YET TO BE RECEIVED. WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED ON (B)(6) 2020, THE STERILE PROCESSING DEPARTMENT (SPD) NOTICED THAT THE TWO (2) DRIVE SHAFTS, A UNIVERSAL CHUCK AND A DRIVING CAP WERE BROKEN. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) DRIVE SHAFT-MINIMUM 520MM LENGTH-FOR USE WITH RIA. THIS IS REPORT 4 OF 4 FOR COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
143561 DRIVE SHAFT-MINIMUM 520MM LENGTH-FOR USE WITH RIA REAMER HTO WRIGHTS LANE SYNTHES USA PRODUCTS LLC 314.743 14382-01 10886982189042

Patients

Seq Age Sex Outcome Treatment
1