FDA Adverse Event Injury Summary report: N

EQUINOXE

MDR report key: 9681199 · Received February 7, 2020

Report

Report Number
1038671-2020-00035
Event Type
Injury
Date Received
February 7, 2020
Date of Event
January 31, 2020
Report Date
December 18, 2020
Manufacturer
EXACTECH, INC.
Product Code
KWT
UDI-DI
10885862086723
PMA / PMN Number
K063569
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

SECTION H10: (D4) SERIAL NUMBER: (B)(6), EXPIRATION DATE: 10-MAR-2024. (H3) THE REVISION REPORTED IN CASE: (B)(4) WAS LIKELY THE RESULT OF EITHER INCOMPLETE SEATING OF THE HUMERAL LINER DURING IMPLANTATION, FORCEFUL CONTACT BETWEEN THE LINER AND GLENOID BONE (SCAPULAR NOTCHING), COMPLETE WEAR OF THE LINER, PATIENT CONDITIONS, OR A COMBINATION OF THE FOUR WHICH LED TO THE LINER DISASSOCIATING FROM THE HUMERAL TRAY. HOWEVER, THIS CANNOT BE CONFIRMED AS THE DEVICES WERE NOT AVAILABLE FOR EVALUATION. (H4) DEVICE MANUFACTURE DATE: 12-MAR-2019. SECTION H11: THE FOLLOWING SECTIONS HAVE CORRECTED INFORMATION: CONCOMITANT DEVICE: 308-10-00, 5367407 - MED PROX BODY +0, 308-15-01, 5535811 - TAPER LOCKING SCREW 0, 320-02-42, 5610472 - RS EXPANDED GLENOSPHERE 42MM, +4MM OFFSET, 320-10-10, 5808447 - EQUINOXE REVERSE TRAY ADAPTER PLATE TRAY +10, 320-15-05, 6216603 - EQ REV LOCKING SCREW, 320-20-00, 6240625 - EQ REVERSE TORQUE DEFINING SCREW KIT.

Additional Manufacturer Narrative · 1

PENDING EVALUATION. CONCOMITANT DEVICE(S): EQUINOXE REVERSE TRAY ADAPTER PLATE TRAY +10 (CN: 320-10-10; SN: NOT REPORTED).

Description of Event or Problem · 1

AS REPORTED, THIS (B)(6) Y/O MALE PATIENT, WEIGHT (B)(6) LBS WAS INITIALLY IMPLANTED WITH A RIGHT TSA ON (B)(6) 2019. A REVISION WAS COMPLETED DUE TO A 42MM +2.5 CONSTRAINED HUMERAL LINER DISASSOCIATING FROM A +10MM HUMERAL ADAPTOR TRAY. IMPLANTS GO TO PATHOLOGY AFTER CASE, WORKING TO RETRIEVE THEM. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
143679 EQUINOXE REVERSE 42MM HUMERAL CONST LINER +2.5 KWT EXACTECH, INC. UNK UNK 10885862086723

Patients

Seq Age Sex Outcome Treatment
1 58 YR Hospitalization| R SEE H10.