HEATER-COOLER SYSTEM 3T
Report
- Report Number
- 9611109-2020-00062
- Event Type
- Injury
- Date Received
- February 7, 2020
- Report Date
- February 6, 2020
- Manufacturer
- LIVANOVA DEUTSCHLAND GMBH
- Product Code
- DWC
- Removal / Correction Number
- Z-2076/2081-2015
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
H.10: FURTHER INFORMATION HAS BEEN REQUESTED TO THE HOSPITAL IN ORDER TO CONFIRM/DENY THE NEWS. DESPITE THE ATTEMPTS, THE HOSPITALS DID NOT PROVIDE ANY FEEDBACK TO LIVANOVA. THUS, NO FURTHER INVESTIGATION IS POSSIBLE.
SEE INITIAL REPORT.
PATIENT INFORMATION WAS NOT PROVIDED. SERIAL NUMBER IS UNKNOWN. THIS INFORMATION WILL BE PROVIDED IN A SUPPLEMENTAL REPORT IF MADE AVAILABLE. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K052601). AS THE SERIAL NUMBER IS UNKNOWN, THE DEVICE MANUFACTURE DATE COULD NOT BE DETERMINED. THIS INFORMATION WILL BE PROVIDED IN A SUPPLEMENTAL REPORT IF MADE AVAILABLE. LIVANOVA (B)(4) IMPLEMENTED A FIELD SAFETY NOTICE FOR DISINFECTION AND CLEANING OF HEATER-COOLER DEVICES. THE Z NUMBER IS Z-2076/2081-2015. LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T . THE INCIDENT OCCURRED IN (B)(6). IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.
THROUGH MEDIA MONITORING LIVANOVA LEARNED OF AN INFECTION OF A PATIENT WITH M.CHIMAERA. IN THE PAPER ARTICLE IT IS STATED THAT THE PATIENT HAD STARTED TO FEEL ILL DURING 2019 SHOWING INCREASING WEAKNESS, A HIGH WEIGHT LOSS, SOME FEVER. THE INFECTED PATIENT, HOSPITALIZED IN THE LAST FEW MONTHS AT (B)(6), HAS NOW BEEN DISCHARGED TO CONTINUE ANTIBIOTIC THERAPY AT HOME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 143943 | HEATER-COOLER SYSTEM 3T | CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS | DWC | LIVANOVA DEUTSCHLAND GMBH | 16-02-80 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |