FDA Adverse Event Malfunction Summary report: N

BD NOKOR¿ FILTER NEEDLE

MDR report key: 9678545 · Received February 6, 2020

Report

Report Number
1911916-2020-00106
Event Type
Malfunction
Date Received
February 6, 2020
Date of Event
January 31, 2020
Report Date
February 3, 2020
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
GAA
UDI-DI
30382903052005
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. INVESTIGATION SUMMARY: SINCE NO SAMPLES DISPLAYING THE CONDITION REPORTED ARE AVAILABLE FOR EXAMINATION, WE WERE UNABLE TO FULLY INVESTIGATE THIS INCIDENT. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED WITH ZERO DEFECTS FOUND. NO QUALITY NOTIFICATIONS WERE WRITTEN FOR THIS BATCH, NOR FOR THE ASSOCIATED ASSEMBLY BATCHES. INVESTIGATION CONCLUSION: BASED ON NO SAMPLE, THE INVESTIGATION CONCLUDED, BD WAS NOT ABLE TO VERIFY THE INDICATED FAILURE. THIS IS THE 2ND COMPLAINT FOR LOT # 8052947 FOR THIS TYPE OF DEFECT OR SYMPTOM. PREVIOUS COMPLAINT (B)(4). THERE WAS NO DOCUMENTATION FOR THIS TYPE OF DEFECTS DURING THE ENTIRE PRODUCTION RUN OF THIS BATCH. ROOT CAUSE DESCRIPTION: NO ROOT CAUSE CAN BE DETERMINED AS NO SAMPLES WERE RECEIVED. RATIONALE: CAPA NOT REQUIRED AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BD NOKOR¿ FILTER NEEDLE BROKE DURING USE. THIS COMPLAINT WAS CREATED TO CAPTURE THE 3RD OF 5 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "BROKEN FILTER NEEDLE."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
142909 BD NOKOR¿ FILTER NEEDLE MANUAL SURGICAL INSTRUMENT FOR GENERAL USE GAA BD MEDICAL (BD WEST) MEDICAL SURGICAL 305200 8052947 30382903052005

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other