FDA Adverse Event Malfunction Summary report: N

SPIROS®, CLOSED MALE LUER W/RED CAP, 10 UNITS

MDR report key: 9673100 · Received February 5, 2020

Report

Report Number
9617594-2020-00037
Event Type
Malfunction
Date Received
February 5, 2020
Date of Event
January 16, 2020
Report Date
January 16, 2020
Manufacturer
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
Product Code
FPA
UDI-DI
00840619042769
PMA / PMN Number
K070532
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A USED CH2000SC-10 SPINNING SPIROS WAS RETURNED CONNECTED AT THE DISTAL END OF AN ALARIS PUMP SET TO A MALE SPIN LUER. THE SPIN FEATURE WAS ACTIVATED IN THE ROTATIONAL DIRECTION AS DESIGNED AND SPINNING FREELY WHEN RECEIVED FOR INVESTIGATION. WHEN EVALUATED THE SPIN COLLAR OF THE ALARIS MALE SPIN LUER WAS SECURELY ATTACHED TO THE FEMALE LUER OF THE CH2000SC-10 SPINNING SPIROS. THERE WAS NO DAMAGE TO THE USED CH2000SC-10 SPINNING SPIROS THAT WOULD HAVE RESULTED IN A PREMATURE DISCONNECT. THE COMPLAINT WAS UNABLE TO BE CONFIRMED. A DEVICE HISTORY REVIEW FOR LOT# 4120021 AND RELEVANT COMMODITIES WERE REVIEWED, AND NO NON-CONFORMANCES WERE FOUND THAT WOULD HAVE CONTRIBUTED TO THE REPORTED COMPLAINT.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A SPIROS, CLOSED MALE LUER W/RED CAP THAT WHEN YOU ATTACH THE SPIROS TO THE END OF A PRIMARY TUBING OR NITRO TUBING THAT IT LOCKS BUT YOU CAN UNLOCK IT. THE CUSTOMER DID CONFIRM THAT IT DOES NOT LEAK; HOWEVER, THEY HAVE A NURSE WHO PULLED IT OFF BY ACCIDENT AND THE CHEMO WERE EVERYWHERE. THE CUSTOMER STATED ANYTIME A SPIROS IS REMOVED FROM THE TUBING, THERE IS A SMALL AMOUNT OF EXPOSURE. ADDITIONALLY, IN THE EVENT OF A HAZARDOUS SPILL, THE HAZARDOUS SPILL PROTOCOL IS ALWAYS FOLLOWED. THERE IS PATIENT INVOLVEMENT AND A DELAY IN THERAPY, BUT THERE WAS NO HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
132606 SPIROS®, CLOSED MALE LUER W/RED CAP, 10 UNITS SET, ADMINISTRATION, INTRAVASCULAR FPA ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 4120021 00840619042769

Patients

Seq Age Sex Outcome Treatment
1