FDA Adverse Event Injury Summary report: N

BARDEX® LUBRI-SIL® I.C. ALL-SILICONE TEMP SENSING FOLEY CATHETER

MDR report key: 9671961 · Received February 5, 2020

Report

Report Number
1018233-2020-00743
Event Type
Injury
Date Received
February 5, 2020
Report Date
February 21, 2020
Manufacturer
C.R. BARD, INC. (COVINGTON) -1018233
Product Code
MJC
PMA / PMN Number
K070582
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS NOT RETURNED FOR EVALUATION. A POTENTIAL ROOT CAUSE COULD BE DUE TO A "KINKED LUMEN". THE DEVICE HISTORY RECORD WAS REVIEWED AND FOUND NOTHING THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. THE INSTRUCTIONS FOR USE WERE FOUND ADEQUATE AND STATE THE FOLLOWING: ¿DEVELOPMENT RELEASED BARD® EZ-LOK® SAMPLING PORT (INDICATED BY THE BLUE STEM IN THE PORT) ACCEPTS LUER LOCK (FIG. 1A) OR SLIP TIP SYRINGES (FIG. 1B). 1. OCCLUDE DRAINAGE TUBING A MINIMUM OF 3 INCHES BELOW THE SAMPLING PORT BY KINKING THE TUBING UNTIL URINE IS VISIBLE UNDER THE ACCESS SITE. 2. SWAB SURFACE OF BARD® EZ-LOK® SAMPLING PORT WITH ANTISEPTIC WIPE. (FIG. 2) 3. USING ASEPTIC TECHNIQUE, POSITION THE SYRINGE IN THE CENTER OF THE SAMPLING PORT. THE SYRINGE SHOULD BE HELD PERPENDICULAR TO THE SURFACE OF THE SAMPLING PORT (AT APPROXIMATELY 80 - 100 DEGREE ANGLE). PRESS THE SYRINGE FIRMLY AND TWIST GENTLY TO ACCESS THE SAMPLING PORT. (FIG. 3) 4. SLOWLY ASPIRATE URINE SAMPLE INTO SYRINGE AND REMOVE SYRINGE FROM SAMPLE PORT. (FIG. 4) 5. UNKINK TUBING AND TRANSFER URINE SPECIMEN INTO SPECIMEN CUP OR FOLLOW HOSPITAL PROTOCOL. DISCARD SYRINGE ACCORDING TO HOSPITAL PROTOCOL. 6. FOLLOW ESTABLISHED HOSPITAL PROTOCOL FOR SPECIMEN LABELING AND TRANSPORT TO LAB. BARD, EZ-LOK AND STATLOCK ARE TRADEMARKS AND/OR REGISTERED TRADEMARKS OF C. R. BARD, INC. ©2018 C. R. BARD, INC. ALL RIGHTS RESERVED. WWW.BARDMEDICAL.COM MANUFACTURER: C. R. BARD, INC. 8195 INDUSTRIAL BLVD. COVINGTON, GA 30014 USA 1.800.526.4455 STERILIZED USING ETHYLENE OXIDE STERILE UNLESS PACKAGE IS OPENED OR DAMAGED, EXCEPT FOR ANY INDIVIDUALLY PACKAGED COMPONENTS WITHIN THE TRAY WHICH ARE NOT LABELED AS STERILE. THESE COMPONENTS ARE NOT TERMINALLY STERILIZED. CAUTION: FEDERAL (U.S.A.) LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN. SINGLE USE ONLY. DO NOT RE-STERILIZE. FOR UROLOGICAL USE ONLY. WARNING: ON CATHETER, DO NOT USE OINTMENTS OR LUBRICANTS HAVING A PETROLATUM BASE. THEY WILL DAMAGE THE CATHETER AND MAY CAUSE BALLOON TO BURST. WARNING: AFTER USE, THIS PRODUCT MAY BE A POTENTIAL BIOHAZARD. HANDLE AND DISPOSE OF IN ACCORDANCE WITH ACCEPTED MEDICAL PRACTICES AND APPLICABLE LOCAL, STATE AND FEDERAL LAWS AND REGULATIONS. VISUALLY INSPECT THE PRODUCT FOR ANY IMPERFECTIONS OR SURFACE DETERIORATION PRIOR TO USE. IF PACKAGE IS OPENED OR IF ANY IMPERFECTION OR SURFACE DETERIORATION IS OBSERVED, DO NOT USE. PLEASE CONSULT PRODUCT LABEL AND INSERT FOR ANY INDICATIONS, CONTRAINDICATIONS, HAZARDS, WARNINGS, CAUTIONS AND DIRECTIONS FOR USE. UROLOGY CDC GUIDELINES FOR APPROPRIATE INDICATIONS FOR INDWELLING URETHRAL CATHETER USE DIRECTIONS FOR USE/PATIENT EDUCATION INFORMATION INSERT FOLEY CATHETERS ONLY FOR APPROPRIATE INDICATIONS AND LEAVE IN PLACE ONLY AS LONG AS NEEDED THIS CATHETER IS INTENDED FOR USE IN THE DRAINAGE AND/OR COLLECTION AND/OR MEASUREMENT OF URINE.* PROPER TECHNIQUES FOR URINARY CATHETER MAINTENANCE SECURE THE FOLEY CATHETER. USE THE STATLOCK® FOLEY STABILIZATION DEVICE IF PROVIDED MAINTAIN A CLOSED DRAINAGE SYSTEM BY UTILIZING PRE-CONNECTED, SEALED CATHETER-TUBING JUNCTIONS MAINTAIN UNOBSTRUCTED URINE FLOW AND KEEP THE CATHETER AND COLLECTION TUBE FREE FROM KINKING KEEP THE COLLECTION BAG BELOW THE LEVEL OF THE BLADDER OR HIPS AT ALL TIMES EMPTY THE COLLECTION BAG REGULARLY USING A SEPARATE, CLEAN COLLECTION CONTAINER FOR EACH PATIENT PROPER TECHNIQUES FOR URINARY CATHETER INSERTION PERFORM HAND HYGIENE IMMEDIATELY BEFORE AND AFTER INSERTION INSERT URINARY CATHETERS USING ASEPTIC TECHNIQUE AND STERILE EQUIPMENT USE THE SMALLEST FOLEY CATHETER POSSIBLE, CONSISTENT WITH GOOD DRAINAGE DOCUMENT THE INDICATIONS FOR CATHETER INSERTION, DATE AND TIME OF CATHETER INSERTION, INDIVIDUAL WHO INSERTED CATHETER, AND DATE AND TIME OF CATHETER REMOVAL IN THE PATIENT RECORD * GENERALLY, DRAINAGE IS ACCOMPLISHED BY INSERTING THE CATHETER THROUGH THE URETHRA AND INTO THE BLADDER. HOWEVER, DRAINAGE IS SOMETIMES ACCOMPLISHED BY SUPRAPUBIC OR OTHER PLACEMENT OF THE CATHETER, SUCH AS NEPHROSTOMY TRACT. ¿ TO ASSIST IN HEALING OF OPEN SACRAL OR PERINEAL WOUNDS ¿ PATIENT REQUIRES PROLONGED IMMOBILIZATION ¿ TO IMPROVE COMFORT FOR END OF LIFE CARE ¿ PATIENT HAS ACUTE URINARY RETENTION OR BLADDER OUTLET OBSTRUCTION ¿ NEED FOR ACCURATE URINE OUTPUT MEASUREMENTS ¿ USE FOR SELECTED SURGICAL PROCEDURES UROLOGY" H11:SECTION A THROUGH F - THE INFORMATION PROVIDED BY BD REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD. H3 OTHER TEXT : THE DEVICE WAS NOT RETURNED.

Additional Manufacturer Narrative · 0

THE DEVICE WAS NOT RETURNED FOR EVALUATION. THE POTENTIAL ROOT CAUSE FOR THIS FAILURE MODE COULD BE USER RELATED (EXAMPLE: OVER ASPIRATED, INCORRECT SYRINGE)/ COLLAPSE LUMEN/SAC CLOSE EYE/ VALVE DAMAGE. THE LOT NUMBER IS UNKNOWN THEREFORE THE DEVICE HISTORY RECORD COULD NOT BE REVIEWED. THE INSTRUCTIONS FOR USE WERE FOUND ADEQUATE AND STATE THE FOLLOWING. H11:SECTION A THROUGH F - THE INFORMATION PROVIDED BY BD REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD. H3 OTHER TEXT : THE DEVICE WAS NOT RETURNED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE FOLEY CATHETER DID NOT DEFLATE AND THE BALLOON HAD TO BE POPPED WITH A GUIDEWIRE BY THE UROLOGIST. THERE WAS NOT IMMEDIATE SUCCESS IN DEFLATING THE BALLOON; HOWEVER, THE BALLOON DEFLATED SHORTLY THERE AFTER.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE FOLEY CATHETER DID NOT DEFLATE AND THE BALLOON HAD TO BE POPPED WITH A GUIDEWIRE BY THE UROLOGIST. THERE WAS NOT IMMEDIATE SUCCESS IN DEFLATING THE BALLOON; HOWEVER, THE BALLOON DEFLATED SHORTLY THERE AFTER.

Additional Manufacturer Narrative · 1

THE INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. THE INFORMATION PROVIDED BY BD REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT/REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD. THE DEVICE WAS NOT RETURNED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE FOLEY CATHETER DID NOT DEFLATE AND THAT THE BALLOON HAD TO BE POPPED WITH A GUIDEWIRE BY THE UROLOGIST.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
136451 BARDEX® LUBRI-SIL® I.C. ALL-SILICONE TEMP SENSING FOLEY CATHETER FOLEY CATHETER MJC C.R. BARD, INC. (COVINGTON) -1018233 UNK

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention