FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES

MDR report key: 9666275 · Received February 4, 2020

Report

Report Number
1024879-2020-00054
Event Type
Malfunction
Date Received
February 4, 2020
Date of Event
January 13, 2020
Report Date
February 26, 2020
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
JKA
UDI-DI
50382903679862
PMA / PMN Number
BK050036
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: BD HAD NOT RECEIVED SAMPLES, BUT PHOTOS WERE PROVIDED BY THE CUSTOMER FACILITY FOR INVESTIGATION. THE PHOTOS WERE EVALUATED AND THE CUSTOMER'S INDICATED FAILURE MODE FOR LABEL LIFT WITH THE INCIDENT LOT WAS OBSERVED. A REVIEW OF THE MANUFACTURING RECORDS WAS COMPLETED FOR THE INCIDENT LOT AND NO ISSUES WERE IDENTIFIED. BD HAS INITIATED FURTHER INVESTIGATION RELATING TO THIS ISSUE THROUGH CAPA#1064141 AND POTENTIAL CAUSES HAVE BEEN IDENTIFIED. AS A RESULT, CORRECTIVE ACTIONS HAVE BEEN ESTABLISHED AND ARE IN THE PROCESS OF BEING IMPLEMENTED. H3 OTHER TEXT : SEE H.10

Description of Event or Problem · 0

IT WAS REPORTED BEFORE USE THE BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES HAD LABEL LIFT OFF/PARTIAL PEEL OFF. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "BEFORE USE THIS BATCH, THE CUSTOMER FOUND MANY TUBES HAVE LABEL LIFT ISSUE. AFFECT QTY: (B)(4) EA."

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED BEFORE USE THE BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES HAD LABEL LIFT OFF/PARTIAL PEEL OFF. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "BEFORE USE THIS BATCH, THE CUSTOMER FOUND MANY TUBES HAVE LABEL LIFT ISSUE. AFFECT QTY: 3000EA."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
132079 BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO., (BD) 367986 9074743 50382903679862

Patients

Seq Age Sex Outcome Treatment
1 Other