OMNIPOD INSULIN PUMP
Report
- Report Number
- 3004464228-2020-01415
- Event Type
- Malfunction
- Date Received
- February 4, 2020
- Date of Event
- January 11, 2020
- Report Date
- January 13, 2020
- Manufacturer
- INSULET CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K162296
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE WAS RECEIVED WITH THE CANNULA NOT DEPLOYED. NO ISSUES WERE FOUND THAT WOULD RESULT IN A NEEDLE MECHANISM FAILURE AS THE POD WAS DEACTIVATED AFTER THE FIRST PRIMING SEQUENCE ENDED BUT BEFORE THE START OF THE SECOND PRIMING SEQUENCE. THE PRODUCT WAS RETURNED WITH A DIFFERENT LOT NUMBER THAN WAS REPORTED AND NOTED ON THE INITIAL MDR. CORRECTION TO D(4): LOT NUMBER CHANGED FROM UNAVAILABLE TO L44898. EXPIRATION DATE CHANGED FROM UNAVAILABLE TO (B)(6) 2020. MODEL NO CHANGED FROM 14810 TO 19191. CORRECTION TO G(5): PMA/510(K) # CHANGED FROM K122953 TO K162296. UNIQUE IDENTIFIER (UDI) # CHANGED FROM UNAVAILABLE TO (B)(4). CORRECTION TO H(4): DEVICE MFG DATE CHANGED FROM UNAVAILABLE TO (B)(6) 2019.
THE DEVICE WAS NOT RETURNED FOR EVALUATION. WE ARE UNABLE TO CONFIRM THE REPORTED NEEDLE MECHANISM FAILURE OR TO DETERMINE ITS ROOT CAUSE. NO LOT RELEASE RECORDS WERE REVIEWED, AS THE PRODUCT LOT NUMBER WAS NOT PROVIDED. MYLIFE OMNIPOD INSULIN MANAGEMENT SYSTEM ¿ USER GUIDE: MODEL: ENT450, 14518-5C-AW REV E 03/16. USING THE POD 5 / PAGE 53: WARNING: ¿CHECK THE INFUSION SITE AFTER INSERTION TO ENSURE THAT THE CANNULA WAS PROPERLY INSERTED. YOU SHOULD CHECK YOUR BLOOD GLUCOSE 1.5 TO 2 HOURS AFTER EACH POD CHANGE AND CHECK THE INFUSION SITE PERIODICALLY. IF THE CANNULA IS NOT PROPERLY INSERTED, HYPERGLYCEMIA MAY RESULT. VERIFY THERE IS NO WETNESS OR SCENT OF INSULIN, WHERE AS MAY INDICATE THE CANNULA HAS DISLODGED.¿
IT WAS REPORTED THE NEEDLE DID NOT DEPLOY; INDICATING A NEEDLE MECHANISM FAILURE. THE POD WAS NOT WORN AND NO ADVERSE PATIENT IMPACT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 127668 | OMNIPOD INSULIN PUMP | PUMP, INFUSION, INSULIN | LZG | INSULET CORPORATION | 19191 | L44898 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |