FDA Adverse Event Injury Summary report: N

TEOSYAL RHA 3

MDR report key: 9664015 · Received February 4, 2020

Report

Report Number
3005975625-2020-00005
Event Type
Injury
Date Received
February 4, 2020
Date of Event
October 20, 2019
Report Date
February 4, 2020
Manufacturer
TEOXANE SA
Product Code
LMH
PMA / PMN Number
P170002
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

VASCULAR COMPLICATIONS (OCCLUSIONS) ARE WELL KNOWN AND DOCUMENTED ADVERSE REACTIONS AS PART OF HYALURONIC ACID-BASED DERMAL FILLERS INJECTIONS. THEY ARE RELATED TO THE ACCIDENTAL INJECTION OF THE PRODUCT INSIDE OR CLOSE TO A BLOOD VESSEL LEADING TO AN OCCLUSION OR A COMPRESSION AND BLOCKING THE BLOOD FLOW, GENERALLY REFERED TO AS VASCULAR COMPLICATION. IF THE VASCULAR COMPLICATION IS NOT DETECTED/DIAGNOSED AND TREATED TIMELY, IT CAN LEAD TO A SKIN NECROSIS. THE RISK OF VASCULAR COMPLICATION AND SKIN NECROSIS ARE MENTIONED IN THE PRODUCT LABELLING. PLEASE REFER TO THE LITERATURE DATA BELOW. LITERATURE DATA: DELORENZI C. COMPLICATIONS OF INJECTABLE FILLERS, PART 2: VASCULAR COMPLICATIONS. AESTHET SURG J. 2014;34(4):584-600. FUNT D, PAVICIC T. DERMAL FILLERS IN AESTHETICS: AN OVERVIEW OF ADVERSE EVENTS AND TREATMENT APPROACHES. CLIN COSMET INVESTIG DERMATOL 2013;6:295-316. MANAFI A, BARIKBIN B, MANAFI A, HAMEDI ZS, MOGHADAM SA. NASAL ALAR NECROSIS FOLLOWING HYALURONIC ACID INJECTION INTO NASOLABIAL FOLDS: A CASE REPORT. WORLD JOURNAL OF PLASTIC SURGERY. 2015;4(1). HONG, J.Y., ET AL., IMPENDING SKIN NECROSIS AFTER DERMAL FILLER INJECTION: A "GOLDEN TIME" FOR FIRST-AID INTERVENTION. DERMATOL THER, 2017. 30(2). MARUYAMA, S., A HISTOPATHOLOGIC DIAGNOSIS OF VASCULAR OCCLUSION AFTER INJECTION OF HYALURONIC ACID FILLER: FINDINGS OF INTRAVASCULAR FOREIGN BODY AND SKIN NECROSIS. AESTHET SURG J, 2017. 37(9): P. NP102-NP108. SALVAL, A., ET AL., IMPENDING FACIAL SKIN NECROSIS AND OCULAR INVOLVEMENT AFTER DERMAL FILLER INJECTION: A CASE REPORT. AESTHETIC PLAST SURG, 2017. 41(5): P. 1198-1201. WANG, Q., ET AL., VASCULAR COMPLICATIONS AFTER CHIN AUGMENTATION USING HYALURONIC ACID. AESTHETIC PLAST SURG, 2018. 42(2): P. 553-559.

Description of Event or Problem · 1

THE PRESENT EVENT HAPPENED OUTSIDE OF THE U.S., IN (B)(6). ACCORDING TO THE RECEIVED INFORMATION, AFTER BEING INJECTED IN THE LIPS A FEW DAYS AGO, THE PATIENT PRESENTED WITH A SEVERE ADVERSE REACTION, LIKELY TO BA A VASCULAR COMPROMISE IN THE UPPER LIP AREA. ACCORDING TO THE RECEIVED INFORMATION, THE INJECTOR MONITORED THE SITUATION OVER THE WEEKEND, AND SAW HER ON (B)(6) 2019. DURING THIS VISIT, THE INJECTOR CONSTATED THAT THE SYMPTOMS WERE IN LINE WITH A VASCULAR OCCLUSION, AND DID A PATCH TEST AND INJECTED HYALASE 1500 IU IN SEVERAL SITES OF THE AREA, ESPECIALLY UNDER THE VESSEL THAT WAS DAMAGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
129744 TEOSYAL RHA 3 HYALURONIC ACID DERML FILLER GEL LMH TEOXANE SA NOT APPLICABLE TP27L-191321B

Patients

Seq Age Sex Outcome Treatment
1 Other