FDA Adverse Event Malfunction Summary report: N

BD PRECISIONGLIDE¿ NEEDLE

MDR report key: 9662521 · Received February 3, 2020

Report

Report Number
1911916-2020-00088
Event Type
Malfunction
Date Received
February 3, 2020
Date of Event
January 17, 2020
Report Date
January 20, 2020
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMI
UDI-DI
30382903051978
PMA / PMN Number
K021475
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. INVESTIGATION SUMMARY: SINCE NO SAMPLES DISPLAYING THE CONDITION REPORTED ARE AVAILABLE FOR EXAMINATION, WE WERE UNABLE TO FULLY INVESTIGATE THIS INCIDENT. NO ROOT CAUSE CAN BE DETERMINED AS NO SAMPLES WERE RECEIVED. EXAMINATION OF THE PRODUCT INVOLVED MAY PROVIDE CLARIFICATION AS TO THE CAUSE FOR THE REPORTED FAILURE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. BASED ON NO SAMPLE, THE INVESTIGATION CONCLUDED, BD WAS NOT ABLE TO VERIFY THE INDICATED FAILURE. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED WITH ZERO DEFECTS FOUND. NO QUALITY NOTIFICATIONS WERE WRITTEN FOR THIS BATCH, NOR FOR THE ASSOCIATED ASSEMBLY BATCHES. INVESTIGATION CONCLUSION: THIS IS THE 1ST COMPLAINT FOR LOT # 9162539 FOR THIS DEFECT OR SYMPTOM. ROOT CAUSE DESCRIPTION: UNDETERMINED. RATIONALE: CAPA NOT REQUIRED AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO USE FOREIGN MATTER WAS DISCOVERED WITH A BD PRECISIONGLIDE¿ NEEDLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT THERE WAS A BLACK AND SILVERY, POSSIBLY A METAL SHAVING WAS FOUND IN A SEALED 16 GAUGE X1 NEEDLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
126207 BD PRECISIONGLIDE¿ NEEDLE HYPODERMIC SINGLE LUMEN NEEDLE FMI BD MEDICAL (BD WEST) MEDICAL SURGICAL 305197 9162539 30382903051978

Patients

Seq Age Sex Outcome Treatment
1 Other