FDA Adverse Event Malfunction Summary report: N

BD MICRO-FINE PLUS PEN NEEDLE

MDR report key: 9661648 · Received February 3, 2020

Report

Report Number
9616656-2020-00072
Event Type
Malfunction
Date Received
February 3, 2020
Date of Event
January 14, 2020
Report Date
April 10, 2020
Manufacturer
BECTON DICKINSON AND CO.
Product Code
FMI
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY : EIGHTY FOUR SEALED AND ONE OPEN 32G X 4MM PEN NEEDLE SAMPLES WERE RETURNED FROM LOT. NO. 9141565, CAT. NO. 320136. VISUAL EXAMINATION WAS CARRIED OUT ON THE EIGHTY-FOUR SEALED SAMPLES AND NO ISSUES WERE OBSERVED. VISUAL EXAMINATION OF THE OPEN SAMPLE WAS ALSO CARRIED OUT AND EVIDENCE OF A TEARDROP LABEL BEING APPLIED TO THE COVER WAS OBSERVED. AS PER MANUFACTURING, A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS." NO ISSUES WERE OBSERVED WITH THE RETURNED SAMPLES THEREFORE NO ROOT CAUSE CAN BE IDENTIFIED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD MICRO-FINE PLUS¿ PEN NEEDLE WAS MISSING THE INNER SHIELD. THIS WAS DISCOVERED BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: INNER SHIELD WAS MISSING.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD MICRO-FINE PLUS¿ PEN NEEDLE WAS MISSING THE INNER SHIELD. THIS WAS DISCOVERED BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: INNER SHIELD WAS MISSING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
126055 BD MICRO-FINE PLUS PEN NEEDLE PEN NEEDLE FMI BECTON DICKINSON AND CO. 9141565

Patients

Seq Age Sex Outcome Treatment
1 Other