FDA Adverse Event Malfunction Summary report: N

OMNIPOD INSULIN PUMP

MDR report key: 9660426 · Received February 3, 2020

Report

Report Number
3004464228-2020-01280
Event Type
Malfunction
Date Received
February 3, 2020
Date of Event
January 8, 2020
Report Date
January 9, 2020
Manufacturer
INSULET CORPORATION
Product Code
LZG
PMA / PMN Number
K162296
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE HAD THE CANNULA ASSEMBLY FULLY DEPLOYED AND THE NEEDLE COMPLETELY RETRACTED. THE DOWNLOAD DATA INDICATES THAT THE POD COMPLETED BOTH ITS FIRST AND SECOND PRIMING SEQUENCES. DEVICE RAN FOR 56 PULSES. NO DAMAGES OR DEFECTS WERE OBSERVED THAT WOULD RESULT IN A LATE NEEDLE DEPLOYMENT. THE CAUSE OF THE REPORTED EVENT COULD NOT BE DETERMINED.D4 - MODEL NO CHANGED FROM 14810 TO 19191. D4 - LOT NO CHANGED FROM BLANK TO L44894. D4 - CATALOG NO CHANGED FROM ZXY425 TO ZXP425. D4 - EXPIRATION DATE CHANGED FROM BLANK TO 12/10/2020. D4 - UNIQUE IDENTIFIER (UDI) # CHANGED FROM BLANK TO (B)(4). G5 - PMA/510(K) # CHANGED FROM K122953 TO K162296. H4 - DEVICE MFG DATE CHANGED FROM BLANK TO 6/10/2019.

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR EVALUATION. WE ARE UNABLE TO CONFIRM THE REPORTED NEEDLE MECHANISM FAILURE OR TO DETERMINE ITS ROOT CAUSE. NO LOT RELEASE RECORDS WERE REVIEWED, AS THE PRODUCT LOT NUMBER WAS NOT PROVIDED. MYLIFE OMNIPOD INSULIN MANAGEMENT SYSTEM ¿ USER GUIDE: MODEL: ENT450, 14518-5C-AW REV E 03/16. USING THE POD 5 / PAGE 53: WARNING: ¿CHECK THE INFUSION SITE AFTER INSERTION TO ENSURE THAT THE CANNULA WAS PROPERLY INSERTED. YOU SHOULD CHECK YOUR BLOOD GLUCOSE 1.5 TO 2 HOURS AFTER EACH POD CHANGE AND CHECK THE INFUSION SITE PERIODICALLY. IF THE CANNULA IS NOT PROPERLY INSERTED, HYPERGLYCEMIA MAY RESULT. VERIFY THERE IS NO WETNESS OR SCENT OF INSULIN, WHERE AS MAY INDICATE THE CANNULA HAS DISLODGED.¿

Description of Event or Problem · 1

IT WAS REPORTED THAT THE NEEDLE DEPLOYED LATE; INDICATING A NEEDLE MECHANISM FAILURE. THE PATIENT'S BLOOD GLUCOSE LEVELS WERE UNAFFECTED AND THE POD WAS NOT WORN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
124943 OMNIPOD INSULIN PUMP PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 19191 L44894

Patients

Seq Age Sex Outcome Treatment
1