OMNIPOD INSULIN PUMP
Report
- Report Number
- 3004464228-2020-01141
- Event Type
- Malfunction
- Date Received
- January 31, 2020
- Date of Event
- January 5, 2020
- Report Date
- January 6, 2020
- Manufacturer
- INSULET CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K162296
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE WAS RECEIVED WITH THE CANNULA ASSEMBLY FULLY DEPLOYED AND THE NEEDLE COMPLETELY RETRACTED. NEEDLE MECHANISM WAS RESET AND SUCCESSFULLY ABLE TO DEPLOY AS INTENDED DURING INVESTIGATION. NO ISSUES WERE SEEN WITH THE DEVICE OR THE DOWNLOAD DATA. ALTHOUGH NO ISSUES WERE NOTED THAT WOULD RESULT IN A NEEDLE MECHANISM FAILURE, THE CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED.D4 - MODEL NO CHANGED FROM 14810 TO 19191. D4 - LOT NO CHANGED FROM BLANK TO L44894. D4 - CATALOG NO CHANGED FROM ZXY425 TO ZXP425. D4 - EXPIRATION DATE CHANGED FROM BLANK TO 12/10//2020. D4 - UNIQUE IDENTIFIER (UDI) # CHANGED FROM BLANK TO (B)(4). G5 - PMA/510(K) # CHANGED FROM K122953 TO K162296. H4 - DEVICE MFG DATE CHANGED FROM BLANK TO 6/10/2019.
THE DEVICE WAS NOT RETURNED FOR EVALUATION. WE ARE UNABLE TO CONFIRM THE REPORTED NEEDLE MECHANISM FAILURE OR TO DETERMINE ITS ROOT CAUSE. NO LOT RELEASE RECORDS WERE REVIEWED, AS THE PRODUCT LOT NUMBER WAS NOT PROVIDED. MYLIFE OMNIPOD INSULIN MANAGEMENT SYSTEM ¿ USER GUIDE: MODEL: ENT450. 14518-5C-AW REV E 03/16. USING THE POD 5 / PAGE 53. WARNING: CHECK THE INFUSION SITE AFTER INSERTION TO ENSURE THAT THE CANNULA WAS PROPERLY INSERTED. YOU SHOULD CHECK YOUR BLOOD GLUCOSE 1.5 TO 2 HOURS AFTER EACH POD CHANGE AND CHECK THE INFUSION SITE PERIODICALLY. IF THE CANNULA IS NOT PROPERLY INSERTED, HYPERGLYCEMIA MAY RESULT. VERIFY THERE IS NO WETNESS OR SCENT OF INSULIN, WHERE AS MAY INDICATE THE CANNULA HAS DISLODGED.
IT WAS REPORTED THAT THE NEEDLE DEPLOYED LATE; INDICATING A NEEDLE MECHANISM FAILURE. THE PATIENT'S BLOOD GLUCOSE LEVELS WERE NOT AFFECTED AND THE POD WAS NOT WORN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 115867 | OMNIPOD INSULIN PUMP | PUMP, INFUSION, INSULIN | LZG | INSULET CORPORATION | 19191 | L44894 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 120 YR |