FDA Adverse Event Malfunction Summary report: N

VITROS 5,1 FS CHEMISTRY SYSTEM

MDR report key: 965239 · Received December 17, 2007

Report

Report Number
1319681-2007-00381
Event Type
Malfunction
Date Received
December 17, 2007
Date of Event
November 1, 2007
Report Date
November 26, 2007
Manufacturer
ORTHO-CLINICAL DIAGNOSTICS, INC.
Product Code
JJE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER SAW IMPRECISE VITROS AMON RESULTS ON QC SAMPLES. OCD FIELD SERVICE EVALUATED THE VITROS 5,1 FS ANALYZER AND FOUND DAMAGED/WORN INCUBATOR CLIPS IN THE INCUBATOR USED TO PROCESS AMON. THE FIELD ENGINEER REPLACED THE INCUBATOR CAPS AND CLEANED THE INCUBATOR. PRECISION TESTING, POST SERVICE, WAS WITHIN SPECIFICATIONS. OCD FIELD SERVICE PERFORMED REPAIRS TO RETURN DEVICE TO EXPECTED OPERATION. THE ROOT CAUSE WAS DETERMINED TO BE INCUBATOR CONTAMINATION CAUSED BY DAMAGED AND WORN INCUBATOR CLIPS. FIELD SERVICE OF THE DEVICE HAS RESOLVED THE ISSUE. NO PATIENT RESULTS WERE MISREPORTED AND NO PATIENT HARM HAS BEEN ALLEGED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

AN OCD CUSTOMER OBSERVED IMPRECISE QUALITY CONTROL RESULTS FOR THE VITROS AMON ASSAY ON A VITROS 5,1 FS ANALYZER. THE CUSTOMER OBSERVED THE IMPRECISE RESULTS ON 1 NOV 2007, 10 NOV 2007, 19 NOV 2007, AND 20 NOV 2007. THE CUSTOMER DID NOT REPORT THE EVENT TO OCD UNTIL 26 NOV 2007. THERE WAS NO ALLEGATION OF MISREPORTED PATIENT RESULTS AND NO ALLEGATION OF HARM TO ANY PATIENTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITROS 5,1 FS CHEMISTRY SYSTEM CLINICAL CHEMISTRY ANALYZER JJE ORTHO-CLINICAL DIAGNOSTICS, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 YR