FDA Adverse Event Malfunction Summary report: N

MICROPOWER OSCILLATOR SAW, DETACHABLE LEVER

MDR report key: 965011 · Received December 14, 2007

Report

Report Number
1017294-2007-00849
Event Type
Malfunction
Date Received
December 14, 2007
Report Date
November 16, 2007
Manufacturer
CONMED LINVATEC
Product Code
HWE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION RESULTS: THE LOANER SAW WAS RECEIVED FOR INVESTIGATION. DURING TESTING OF THIS UNIT, NO OILY-LIKE RESIDUE LEAKED FROM THE HANDPIECE. A VISUAL EXAMINATION OF THE SAW NOTED AN OILY-LIKE BLACK RESIDUE UNDER THE OSCILLATING HEAD OF THIS HANDPIECE. THE DEVICE WAS SENT TO A THIRD PARTY LAB FOR ANALYSIS OF THE OILY-LIKE RESIDUE ALONG WITH A SAMPLE OF SRI-2 LUBRICANT USED IN THE MANUFACTURE OF THIS PRODUCT. THE LAB ANALYSIS INDICATES THAT THE RESIDUE REMOVED FROM THE OSCILLATING SAW IS COMPOSED OF A FATTY ACID BASED OIL OR COMPOUND WHICH PRECLUDES SRI-2 AS THE SOURCE OF THE UNK RESIDUE. IN ADDITION, THERE ARE NO CUSTOMER REPORTS OF THIS UNIT LEAKING AN OILY-LIKE RESIDUE ON PRIOR USES. THE INSTRUCTION MANUAL FOR THIS HANDPIECE ALONG WITH THE LABELING FOR THIS PRODUCT INFORMS THE USER NOT TO LUBRICATE THIS HANDPIECE. THE CUSTOMER WILL BE CONTACTED WITH ADDITIONAL INFORMATION ON CARE AND MAINTENANCE OF THIS PRODUCT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING USE OF THIS HANDPIECE IN A BUNIONECTOMY PROCEDURE, THE SAW LEAKED A BLACK RESIDUE RESEMBLING OIL INTO THE SURGICAL SITE. THERE WAS NO REPORT OF SERIOUS INJURY OR SURGICAL DELAY WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MICROPOWER OSCILLATOR SAW, DETACHABLE LEVER OSCILLATING SAW, LONANER HWE CONMED LINVATEC NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK YR