FDA Adverse Event Malfunction Summary report: N

BD PLASTIPAK CONCENTRIC LUER LOCK SYRINGE

MDR report key: 9643516 · Received January 29, 2020

Report

Report Number
3003152976-2020-00023
Event Type
Malfunction
Date Received
January 29, 2020
Date of Event
October 4, 2019
Report Date
February 24, 2020
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY. NO PHOTOS OR PHYSICAL SAMPLES THAT DISPLAY THE REPORTED CONDITION WERE AVAILABLE FOR INVESTIGATION. A DEVICE HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT 1802271, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. TEN RETAINED SAMPLES OF LOT 1802271 WERE USED TO CONDUCT A LEAKAGE TEST. THE PRODUCT WAS VISUALLY INSPECTED, NO DEFECTS OR DAMAGE WAS NOTED, AND NO LEAK WAS IDENTIFIED. BASED ON THE AVAILABLE INFORMATION WE ARE NOT ABLE TO DETERMINE A ROOT CAUSE AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEFECT AND DEVICE WILL BE MONITORED BY OUR QUALITY TEAM FOR SIGNS OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE LEAKED PAST THE STOPPER DURING THE INJECTION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: "AFTER SAMPLING THE PRODUCT FROM EACH VIAL, DURING INJECTING, THERE HAS BEEN A LEAK AT THE PAST STOPPER. ACTIONS TAKEN: PACKAGING KEPT TO RAISE THIS COMPLAINT. REDO OF ALL OF THE PREPARATION".

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE LEAKED PAST THE STOPPER DURING THE INJECTION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: "AFTER SAMPLING THE PRODUCT FROM EACH VIAL, DURING INJECTING, THERE HAS BEEN A LEAK AT THE PAST STOPPER. ACTIONS TAKEN: PACKAGING KEPT TO RAISE THIS COMPLAINT. REDO OF ALL OF THE PREPARATION."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
108579 BD PLASTIPAK CONCENTRIC LUER LOCK SYRINGE PISTON SYRINGE FMF BECTON DICKINSON, S.A. 1802271

Patients

Seq Age Sex Outcome Treatment
1 Other