BD PLASTIPAK CONCENTRIC LUER LOCK SYRINGE
Report
- Report Number
- 3003152976-2020-00023
- Event Type
- Malfunction
- Date Received
- January 29, 2020
- Date of Event
- October 4, 2019
- Report Date
- February 24, 2020
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMF
- PMA / PMN Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.6. INVESTIGATION SUMMARY. NO PHOTOS OR PHYSICAL SAMPLES THAT DISPLAY THE REPORTED CONDITION WERE AVAILABLE FOR INVESTIGATION. A DEVICE HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT 1802271, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. TEN RETAINED SAMPLES OF LOT 1802271 WERE USED TO CONDUCT A LEAKAGE TEST. THE PRODUCT WAS VISUALLY INSPECTED, NO DEFECTS OR DAMAGE WAS NOTED, AND NO LEAK WAS IDENTIFIED. BASED ON THE AVAILABLE INFORMATION WE ARE NOT ABLE TO DETERMINE A ROOT CAUSE AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEFECT AND DEVICE WILL BE MONITORED BY OUR QUALITY TEAM FOR SIGNS OF EMERGING TRENDS.
IT WAS REPORTED THAT THE BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE LEAKED PAST THE STOPPER DURING THE INJECTION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: "AFTER SAMPLING THE PRODUCT FROM EACH VIAL, DURING INJECTING, THERE HAS BEEN A LEAK AT THE PAST STOPPER. ACTIONS TAKEN: PACKAGING KEPT TO RAISE THIS COMPLAINT. REDO OF ALL OF THE PREPARATION".
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT THE BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE LEAKED PAST THE STOPPER DURING THE INJECTION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: "AFTER SAMPLING THE PRODUCT FROM EACH VIAL, DURING INJECTING, THERE HAS BEEN A LEAK AT THE PAST STOPPER. ACTIONS TAKEN: PACKAGING KEPT TO RAISE THIS COMPLAINT. REDO OF ALL OF THE PREPARATION."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 108579 | BD PLASTIPAK CONCENTRIC LUER LOCK SYRINGE | PISTON SYRINGE | FMF | BECTON DICKINSON, S.A. | 1802271 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |