ATRAUMATIC GRASPER
MDR report key: 96419
·
Received June 5, 1997
Report
- Report Number
- 1924654-1997-00006
- Event Type
- Malfunction
- Date Received
- June 5, 1997
- Date of Event
- April 11, 1997
- Report Date
- April 15, 1997
- Manufacturer
- ACUFEX MICROSURGICAL, INC.
- Product Code
- GCJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
REFERENCE MFR. COMPLAINT # AF1997-5-6-35. ACTUAL SAMPLE WAS RETURNED AND EXAMINED MICROSCOPICALLY. THE FIXED JAW OF THE DEVICE WAS BROKEN OFF THE GRASPER ASSEMBLY AND THE MECHANISM THAT HOLDS THE JAW WAS MISSING. INVESTIGATION CONCLUDED NO SIGNS OF ABUSE WERE PRESENT. THE DEVICE MOST PROBABLY BROKE DUE TO OVERLOADING. A REVIEW OF THE LOT HISTORY FOUND THAT AT THE TIME OF MFR THE DEVICE MET ALL SPECIFICATIONS.
Description of Event or Problem · 1
CUSTOMER REPORTS, DAVIS AND GECK GRASPER BEING USED TO GRASP THE GALL BLADDER IN A LAPAROSCOPIC PROCEDURE. THE TIP BROKE OFF. X-RAY WAS REQUIRED TO LOCATE THE BROKEN TIP. THE TIP WAS RETRIEVED LAPAROSCOPICALLY WITH NO INJURY TO THE PATIENT. THE SURGERY WAS DELAYED FOR APPROXIMATELY THIRTY MINUTES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ATRAUMATIC GRASPER | GRASPER | GCJ | ACUFEX MICROSURGICAL, INC. | NA | 888053 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR |