FDA Adverse Event Malfunction Summary report: N

BD AUTOSHIELD DUO SAFETY PEN NEEDLE

MDR report key: 9635887 · Received January 28, 2020

Report

Report Number
9616656-2020-00039
Event Type
Malfunction
Date Received
January 28, 2020
Date of Event
January 7, 2020
Report Date
February 26, 2020
Manufacturer
BECTON DICKINSON AND CO.
Product Code
FMI
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: TWO OPEN 30G X 5MM SAFETY PEN NEEDLE SAMPLES WERE RETURNED FROM LOT NO. 9141984, CAT. NO. 329605 ALONG WITH ONE OPEN 30G X 5MM SAFETY PEN NEEDLE SAMPLE FROM LOT NO. 9051755, CAT. NO. 329605. A CLOG TEST AS PER TP700483 WAS CARRIED OUT ON ALL THREE SAMPLES AND NO ISSUES WERE OBSERVED. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS. NO ISSUES WERE OBSERVED WITH THE RETURNED SAMPLES THEREFORE NO ROOT CAUSE CAN BE IDENTIFIED. H3 OTHER TEXT : SEE H.10

Description of Event or Problem · 0

IT WAS REPORTED THAT BD AUTOSHIELD¿ DUO SAFETY PEN NEEDLE WAS CLOGGED. THIS WAS DISCOVERED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: COMPLAINT: PROBLEMS WITH ACCESSIBILITY, THE NEEDLE HAS BEEN SAVED, CLOGGED NEEDLES.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD AUTOSHIELD¿ DUO SAFETY PEN NEEDLE WAS CLOGGED. THIS WAS DISCOVERED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: COMPLAINT: PROBLEMS WITH ACCESSIBILITY, THE NEEDLE HAS BEEN SAVED, CLOGGED NEEDLES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
103232 BD AUTOSHIELD DUO SAFETY PEN NEEDLE PEN NEEDLE FMI BECTON DICKINSON AND CO. 9051755

Patients

Seq Age Sex Outcome Treatment
1 Other