BD AUTOSHIELD DUO SAFETY PEN NEEDLE
Report
- Report Number
- 9616656-2020-00039
- Event Type
- Malfunction
- Date Received
- January 28, 2020
- Date of Event
- January 7, 2020
- Report Date
- February 26, 2020
- Manufacturer
- BECTON DICKINSON AND CO.
- Product Code
- FMI
- PMA / PMN Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.6. INVESTIGATION: TWO OPEN 30G X 5MM SAFETY PEN NEEDLE SAMPLES WERE RETURNED FROM LOT NO. 9141984, CAT. NO. 329605 ALONG WITH ONE OPEN 30G X 5MM SAFETY PEN NEEDLE SAMPLE FROM LOT NO. 9051755, CAT. NO. 329605. A CLOG TEST AS PER TP700483 WAS CARRIED OUT ON ALL THREE SAMPLES AND NO ISSUES WERE OBSERVED. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS. NO ISSUES WERE OBSERVED WITH THE RETURNED SAMPLES THEREFORE NO ROOT CAUSE CAN BE IDENTIFIED. H3 OTHER TEXT : SEE H.10
IT WAS REPORTED THAT BD AUTOSHIELD¿ DUO SAFETY PEN NEEDLE WAS CLOGGED. THIS WAS DISCOVERED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: COMPLAINT: PROBLEMS WITH ACCESSIBILITY, THE NEEDLE HAS BEEN SAVED, CLOGGED NEEDLES.
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT BD AUTOSHIELD¿ DUO SAFETY PEN NEEDLE WAS CLOGGED. THIS WAS DISCOVERED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: COMPLAINT: PROBLEMS WITH ACCESSIBILITY, THE NEEDLE HAS BEEN SAVED, CLOGGED NEEDLES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 103232 | BD AUTOSHIELD DUO SAFETY PEN NEEDLE | PEN NEEDLE | FMI | BECTON DICKINSON AND CO. | 9051755 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |