FDA Adverse Event Malfunction Summary report: N

SAMSUNG

MDR report key: 9617491 · Received January 22, 2020

Report

Report Number
1125873-2020-00001
Event Type
Malfunction
Date Received
January 22, 2020
Report Date
January 2, 2020
Manufacturer
VISICU, INC - DUP
Product Code
MSX
UDI-DI
00884838084452
PMA / PMN Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION AND OBJECTIVE EVIDENCE REVEALED NO BURNING, MELTING OR CHARRING OF THE ADAPTOR OR CORD. THE ADAPTOR WAS IDENTIFIED AS AN UNAPPROVED/SUPPORTED/SUPPLIED ADAPTOR USED BY THE COMPLAINANT.

Description of Event or Problem · 1

IT WAS REPORTED ON (B)(6) 2019 THAT THE CORD TO THE TABLET WAS ON FIRE. NURSE ASKED PATIENT IF THERE WERE ACTIVE SPARKS OR FIRE. PATIENT REPORTED SHE WAS UNSURE. NURSE CONFIRMED THERE WAS NO INJURY TO THE PATIENT. INSTALLER RETURNED TO THE PATIENT HOME ON (B)(6) 2020 TO RETRIEVE THE CORD AND TABLET AND SUBSEQUENTLY RETURNED THE ENTIRE KIT INCLUDING THE ECAD BP, SPO2, AND THE SCALE WHICH IS THE ENTIRE KIT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
82671 SAMSUNG N/A MSX VISICU, INC - DUP SAMSUNG TAB E SM-T377A NA 00884838084452

Patients

Seq Age Sex Outcome Treatment
1 53 YR Female Other