FDA Adverse Event
Malfunction
Summary report: N
SAMSUNG
MDR report key: 9617491
·
Received January 22, 2020
Report
- Report Number
- 1125873-2020-00001
- Event Type
- Malfunction
- Date Received
- January 22, 2020
- Report Date
- January 2, 2020
- Manufacturer
- VISICU, INC - DUP
- Product Code
- MSX
- UDI-DI
- 00884838084452
- PMA / PMN Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 1
INVESTIGATION AND OBJECTIVE EVIDENCE REVEALED NO BURNING, MELTING OR CHARRING OF THE ADAPTOR OR CORD. THE ADAPTOR WAS IDENTIFIED AS AN UNAPPROVED/SUPPORTED/SUPPLIED ADAPTOR USED BY THE COMPLAINANT.
Description of Event or Problem · 1
IT WAS REPORTED ON (B)(6) 2019 THAT THE CORD TO THE TABLET WAS ON FIRE. NURSE ASKED PATIENT IF THERE WERE ACTIVE SPARKS OR FIRE. PATIENT REPORTED SHE WAS UNSURE. NURSE CONFIRMED THERE WAS NO INJURY TO THE PATIENT. INSTALLER RETURNED TO THE PATIENT HOME ON (B)(6) 2020 TO RETRIEVE THE CORD AND TABLET AND SUBSEQUENTLY RETURNED THE ENTIRE KIT INCLUDING THE ECAD BP, SPO2, AND THE SCALE WHICH IS THE ENTIRE KIT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 82671 | SAMSUNG | N/A | MSX | VISICU, INC - DUP | SAMSUNG TAB E SM-T377A | NA | 00884838084452 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR | Female | Other |