FDA Adverse Event Malfunction Summary report: N

METAL HANDLE OFFSET CUP IMPACTOR

MDR report key: 9615979 · Received January 22, 2020

Report

Report Number
3004976965-2019-00049
Event Type
Malfunction
Date Received
January 22, 2020
Date of Event
December 17, 2019
Report Date
December 24, 2019
Manufacturer
VIANT MEDICAL, LLC
Product Code
HWA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINT SAMPLE WAS RECEIVED AT VIANT FOR EVALUATION AND THE REPORTED EVENT WAS CONFIRMED. THE WELD ADJOINING THE RATCHET HOUSING AND THE OFFSET CUP IMPACTOR (OCI) WAS FRACTURED ALL AROUND AND THE RATCHET HOUSING WAS REMOVED FROM THE OCI BODY. THERE WERE SEVERAL DEFORMITIES ON THE RATCHET HOUSING, AS WELL AS IN THE SURROUNDING AREA OF THE RATCHET ASSEMBLY, WHICH ARE INCONSISTENT WITH INTENDED USE. THERE WERE SOME SIGNS OF WEAR IN THE FORM OF SCRATCHES, NICKS AND GOUGES OBSERVED THROUGHOUT THE COMPLAINT SAMPLE; THERE WERE SOME GOUGES ON THE METAL HANDLE THAT ARE NOT CONSISTENT WITH INTENDED USE AS THE OCI IS INTENDED TO BE IMPACTED ONLY ON THE IMPACTION PLATE; THE OCI BODY, NEAR THE 2 LARGE PINS THAT ATTACH INTO THE OCI BODY, WAS DENTED ON ONE (1) SIDE, LIKELY FROM A LARGE IMPACTION IN THIS AREA, WHICH IS NOT INTENDED; THE RETURNED COMPLAINT SAMPLE WAS ETCHED PER APPLICABLE DRAWINGS. REVIEW OF PRODUCTION RECORDS DID NOT REVEAL ANY DISCREPANCIES. IN CONCLUSION, THE REPORTED EVENT IS CONFIRMED AND IS ATTRIBUTED TO UNINTENDED USE. NO FURTHER INVESTIGATION WITH REGARD TO THIS COMPLAINT IS REQUIRED. REPORTED BY DISTRIBUTOR, DEPUY ORTHOPAEDICS, INC.

Description of Event or Problem · 1

IT WAS REPORTED DURING AN UNKNOWN PATIENT PROCEDURE THAT UPON IMPACTION, THE SPRING LOADED RELEASE MECHANISM DISASSOCIATED FROM THE HANDLE RELEASING TENSION FROM THE DEVICE. NO ADVERSE EVENTS NOR PATIENT CONSEQUENCE WERE REPORTED AS A RESULT OF THE MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
81242 METAL HANDLE OFFSET CUP IMPACTOR IMPACTOR HWA VIANT MEDICAL, LLC 255000115 PC4194837

Patients

Seq Age Sex Outcome Treatment
1