FDA Adverse Event Malfunction Summary report: N

ONCOR EXPRESSION W/COHERENCE THERAPIST

MDR report key: 961551 · Received December 6, 2007

Report

Report Number
2910081-2007-00035
Event Type
Malfunction
Date Received
December 6, 2007
Date of Event
August 10, 2007
Report Date
November 2, 2007
Manufacturer
SIEMENS MEDICAL SOLUTIONS, ONCOLOGY CARE SYSTEMS
Product Code
IYE
PMA / PMN Number
K060226
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

SEE ATTACHMENT TITLED ADDITIONAL MFR'S NARRATIVE MDR 2910081-2007-00035. THIS REPORT SUMMARIZES THIS ISSUE AND INCLUDES OUR RISK ANALYSIS RESULT AND CORRECTIVE ACTION.

Description of Event or Problem · 1

IN THE ABUNDANCE OF CAUTION, THIS ISSUE IS BEING REPORTED: SYSTEMS INVOLVED: COHERENCE THERAPIST WORKSPACE (RTT VERSION 2.1) WHEN THE REFERENCE IMAGE HAS BEEN CALIBRATED FOR CENTERING, WHEN THE REFERENCE IMAGE IS USED FOR PT POSITIONING, THE IMAGE WILL SHIFT WHEN THE POSITIONING TOOLS ARE USED. POTENTIAL FOR MISTREATMENT IF THE IMAGE IS SHIFTED SO THAT THE IMAGE ISOCENTER DOES NOT CORRESPOND TO THE MACHINE ISOCENTER. ITS IMPORTANT TO NOTE THAT THE ANOMALY WAS DISCOVERED PRIOR TO TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONCOR EXPRESSION W/COHERENCE THERAPIST ACCELERATOR, LINEAR, MEDICAL IYE SIEMENS MEDICAL SOLUTIONS, ONCOLOGY CARE SYSTEMS NA NA

Patients

Seq Age Sex Outcome Treatment
1 YR