STYLE 15 SILICONE GEL FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2019-20723
- Event Type
- Injury
- Date Received
- January 21, 2020
- Date of Event
- July 15, 2010
- Report Date
- April 22, 2020
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FTR
- PMA / PMN Number
- P020056
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT OF RUPTURE WAS RECEIVED ON (B)(6) 2020 WITH LOT NUMBER 1850503. ANALYSIS OF THE RETURNED DEVICE IDENTIFIED: WEIGHT TO THE SPEC, RED PARTICLES IN THE SHELL, CREASE FOLD AND OPENING ON ANTERIOR. A VISUAL AND MICRO ANALYSIS WAS PERFORMED WHICH IDENTIFIED CREASED SHARP OPENING, STRESS MARKS AND WEAR ABRASION. BASE ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: A OPENING CREASE SHARP ON ANTERIOR ASSESSED AS A FOLD FLAW OPENING. ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: D10, G1, H3, H6.
PATIENT REPORTED UNKNOWN SIDE RUPTURE. DEVICE HAS BEEN EXPLANTED. THIS RECORD IS FOR THE RIGHT SIDE.
ALLERGAN DID NOT SUBMIT THIS MDR WITHIN 30 DAYS OF BECOMING AWARE. RECENT STIMULATED REPORTING RELATED TO 2011068-7/2/19-001-R HAS INCREASED COMPLAINT AND MDR VOLUME. ALLERGAN IS IMPLEMENTING A PLAN TO ADDRESS THE INCREASED VOLUMES. A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.
PATIENT REPORTED UNKNOWN SIDE RUPTURE. DEVICE HAS BEEN EXPLANTED. THIS RECORD IS FOR THE RIGHT SIDE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 76630 | STYLE 15 SILICONE GEL FILLED BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | ALLERGAN (COSTA RICA) | 1850503 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 29 YR | Required Intervention |