ARTICULEZE M HEAD 36MM +5
Report
- Report Number
- 1818910-2020-02233
- Event Type
- Injury
- Date Received
- January 20, 2020
- Date of Event
- November 5, 2019
- Report Date
- January 6, 2020
- Manufacturer
- DEPUY ORTHOPAEDICS INC US
- Product Code
- JDI
- UDI-DI
- 10603295033936
- PMA / PMN Number
- K980513
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISION OF 21 CFR, PART 803. THE REPORT MAY BE BASED ON THE INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. PRODUCT COMPLAINT # : (B)(4). INVESTIGATION SUMMARY : NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. THE INFORMATION RECEIVED WILL BE RETAINED FOR POTENTIAL SERIES INVESTIGATIONS IF TRIGGERED BY TREND ANALYSIS, POST MARKET SURVEILLANCE, OR OTHER EVENTS WITHIN THE QUALITY SYSTEM. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY. DEVICE HISTORY LOT : NULL. DEVICE HISTORY BATCH : NULL. DEVICE HISTORY REVIEW : NULL. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
PRODUCT COMPLAINT # (B)(4). DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT.
ADDITIONAL INFORMATION RECEIVED FOR MEDICAL RECORDS. LAB VALUES FROM (B)(6) 2018: COBALT 7.1 MCG/L AND CHROMIUM 1.6 MCG/L. ON (B)(6) 2019, THE PATIENT HAD A REVISION OF LEFT TOTAL HIP TO ADDRESS METALLOSIS, PAIN, AND ELEVATED METAL ION LEVELS. DURING THE PROCEDURE THE SURGEON NOTED INFLAMMATION, EXTENSIVE TRUNNIONOSIS WITH CORROSION ON THE MORSE TAPER ASPECT OF THE FEMORAL HEAD AND TRUNNION. THE ACETABULAR CUP AND FEMORAL STEM WERE NOTED TO BE WELL FIXED AND NOT REVISED.
(B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
SUBJECT ID: (B)(6). STUDY: (B)(6). CLINICAL NOTIFICATION RECEIVED FOR REVISION OF LEFT TOTAL HIP TO ADDRESS ADVERSE LOCAL TISSUE REACTION. DATE OF IMPLANTATION: (B)(6) 2007, DATE OF REVISION: (B)(6) 2019, (LEFT HIP). LAB: HIGH COBALT BLOOD LEVEL 7.4 ((B)(6) 2018). METAL HEAD AND LINER REVISED. DEPUY COMPONENTS: CATALOG ID: 121701050, LOT: W75FV1012, DESCRIPTION: PINNACLE SHELL 100 SERIES 50MM. CATALOG ID: 121887350, LOT: 2367675, DESCRIPTION: ULTAMET LINER METAL 50MM X 36MM. CATALOG ID: 124603000, LOT: A82DD1000, DESCRIPTION: APEX HOLE ELIM POSITIVE STOP. CATALOG ID: 136552000, LOT: 2316631, DESCRIPTION: ARTICUL/EZE HEAD METAL 36MM +5. CATALOG ID: 3L92511, LOT: 1860160, DESCRIPTION: CORAIL STEM COLLARLESS SZ 11.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 70735 | ARTICULEZE M HEAD 36MM +5 | ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS | JDI | DEPUY ORTHOPAEDICS INC US | 1365-52-000 | 2316631 | 10603295033936 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Required Intervention | APEX HOLE ELIM POSITIVE STOP (124603000/A82DD1000)| PINNACLE 100 ACET CUP 50MM (121701050/W75FV1012) |