FDA Adverse Event Malfunction Summary report: N

CLINAC

MDR report key: 959020 · Received December 4, 2007

Report

Report Number
2914292-2007-00048
Event Type
Malfunction
Date Received
December 4, 2007
Date of Event
November 14, 2007
Report Date
November 14, 2007
Manufacturer
VARIAN MEDICAL SYSTEMS
Product Code
IYE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

METHOD - ACTUAL DEVICE IN INCIDENT WAS EVALUATED BY A VARIAN ENGINEER. VISUAL EXAMINATION WAS MADE. RESULTS: INCORRECT TECHNIQUE/PROCEDURE WAS USED WHEN INSTALLING THE MASTER LINK ON THE GANTRY DRIVE CHAIN. CONCLUSION: DEVICE MAINTENANCE CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN REPAIRED AND RETURNED. THE GANTRY CHAIN MASTER LINK AND THE GANTRY CHAIN HAVE BEEN RETURNED FOR INVESTIGATION BY VARIAN MEDICAL SYSTEMS. THERE WAS NO DAMAGE TO THE SECURING PART OF THE MASTER LINK (LINK PLATE AND SPRING CLIP). IT HAS BEEN DETERMINED THAT THE ROOT CAUSE OF THE GANTRY CHAIN "BREAKING" IS THE MASTER LINK CLIP NOT BEING ENGAGED ALL THE WAY WHEN INSTALLED. TO ALLEVIATE THIS PROBLEM, VARIAN MEDICAL SYSTEMS IS CHANGING THE ASSEMBLY PROCEDURES TO CLARIFY THE PROPER ASSEMBLY AND INSPECTION OF THIS LINK.

Description of Event or Problem · 1

THE CUSTOMER REPORTED, THAT THEY WERE USING THE AUTO-GO-TO FUNCTION TO SET UP CLINAC POSITIONS FOR TREATMENT, WHEN A LOUD SOUND WAS HEARD. THE CUSTOMER NOTICED THAT ALL MOTIONS WERE CONTINUING EXCEPT FOR GANTRY MOVEMENT. THE CUSTOMER THEN CALLED THE VARIAN SERVICE REPRESENTATIVE. UPON INSPECTING THE CLINAC, THE SERVICE REPRESENTATIVE DISCOVERED THAT THE GANTRY CHAIN HAD SEPARATED AT THE MASTER LINK WITH THE GANTRY AT APPROXIMATELY 270 DEGREES. THERE WAS A PT ON THE TREATMENT TABLE, BUT THERE WERE NO INJURIES AS A RESULT OF THIS EVENT. THERE WAS NO TREAT OF A COLLISION, AS THE GANTRY DID NOT MOVE AFTER THE CHAIN BREAK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CLINAC LINEAR ACCELERATOR IYE VARIAN MEDICAL SYSTEMS CL-IX TRILOGY

Patients

Seq Age Sex Outcome Treatment
1 YR