BD PROTECTOR P50
Report
- Report Number
- 3003152976-2019-00917
- Event Type
- Malfunction
- Date Received
- January 14, 2020
- Date of Event
- December 19, 2019
- Report Date
- March 30, 2020
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- ONB
- UDI-DI
- 30382905151058
- PMA / PMN Number
- K123213
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TU
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.6. INVESTIGATION SUMMARY ONE PHOTO SAMPLE WAS PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. UPON VISUALLY INSPECTING THE PHOTO, A GREY PARTICLE WHICH APPEARS TO BE CONSISTENT WITH THE RUBBER STOPPER IS VISIBLE INSIDE THE VIAL. WITHOUT THE ACTUAL SAMPLE TO EVALUATE, WE CANNOT VERIFY THE ORIGIN OF THE PARTICLE. A DEVICE HISTORY REVIEW WAS PERFORMED FOR LOT 1905122, NO DEVIATIONS OR NON-CONFORMANCE'S WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THESE ISSUES. FRAGMENTATION TESTING IS PERFORMED DURING MANUFACTURING TO EVALUATE ANY PARTICULATES GENERATED AFTER TEN ACTIVATIONS. TESTING RESULTS WERE REVIEWED FOR THE REPORTED LOT AND RESULTS WERE FOUND TO BE ACCEPTABLE. FOUR RETAINED SAMPLES OF LOT 1905122 WERE USED FOR ADDITIONAL EVALUATION. FUNCTIONAL TESTING WAS ALSO PERFORMED, PENETRATING THE PROTECTOR WITH A SAMPLE INJECTOR TEN TIMES. IN ALL CASES THE PRODUCT FUNCTIONED AS INTENDED AND NO CORING WAS IDENTIFIED. WHILE PHASEAL NEEDLES ARE DESIGNED TO REDUCE CORING, THERE ARE SEVERAL FACTORS THAT MAY IMPACT CORING TENDENCY. CORING FROM THE MEMBRANE MAY OCCUR DUE TO FRAGMENTATION CAUSED BY MULTIPLE INJECTIONS OR EXCESSIVE WELDING OF THE MEMBRANE. CORING OF THE RUBBER STOPPER MAY RESULT DEPENDING ON THE STOPPER QUALITY, THE DESIGN AND DIMENSIONS OF THE NEEDLES USED, OR IF A POOR CONNECTION OF THE PROTECTOR OCCURS. BASED ON THE QUALITY TEAM'S INVESTIGATION, WE ARE NOT ABLE TO IDENTIFY A DEFINITIVE ROOT CAUSE RELATED TO OUR MANUFACTURING PROCESS AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND DEFECT WILL CONTINUE TO BE MONITORED BY OUR QUALITY TEAM.
IT HAS BEEN REPORTED THAT THE BD¿ PROTECTOR P50 HAS BEEN FOUND EXPERIENCING THREE OCCURRENCES OF CONTAINING FOREIGN MATTER DURING USE. THE FOLLOWING HAS BEEN PROVIDED BY THE INITIAL REPORTER: A PIECE OF VIAL RUBBER FALLING INTO THE VIAL WHEN THE PROTECTOR IS INSERTED INTO THE VIAL.
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT HAS BEEN REPORTED THAT THE BD¿ PROTECTOR P50 HAS BEEN FOUND EXPERIENCING THREE OCCURRENCES OF CONTAINING FOREIGN MATTER DURING USE. THE FOLLOWING HAS BEEN PROVIDED BY THE INITIAL REPORTER: A PIECE OF VIAL RUBBER FALLING INTO THE VIAL WHEN THE PROTECTOR IS INSERTED INTO THE VIAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 52223 | BD PROTECTOR P50 | INTRAVASCULAR ADMINISTRATION SET | ONB | BECTON DICKINSON, S.A. | 515105 | 1905122 | 30382905151058 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |