FDA Adverse Event Other Summary report: N

SURGICAL SPRING CLIPS-DOUBLE SOFT, 6MM

MDR report key: 958321 · Received April 25, 2007

Report

Report Number
2027111-2007-00023
Event Type
Other
Date Received
April 25, 2007
Date of Event
February 26, 2007
Manufacturer
APPLIED MEDICAL RESOURCES
Product Code
DSS
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION: PIECES FROM FOUR OR FIVE DIFFERENT SPRING CLIP UNITS WERE RETURNED AND VISUALLY INSPECTED. THE BODY FEMALE COMPONENT, IS THE ONLY BROKEN COMPONENT RETURNED. EACH OF THE RETURNED BODY FEMALE COMPONENTS SPLIT IN A DIFFERENT WAY; BUT MOST ARE SPLIT BELOW WHERE THE BODY FEMALE SNAP WITH THE FEMALE JAW. CONCLUSION/CORRECTIVE ACTION: THE CAUSE OF THE INCIDENT WAS INVESTIGATED BUT CANNOT BE DETERMINED WITHOUT ADDITIONAL INFO. ENGINEERING WAS UNABLE TO DUPLICATE THE REPORTED INCIDENT USING A FULLY ACTUATED MODEL A3218 CLIP APPLICATOR. THE TYPE OF CLIP APPLICATOR USED, THE AMOUNT OF FORCE APPLIED AND OTHER FACTORS ARE NEEDED TO UNDERSTAND THE CONDITION THE CLIP WAS IN.

Description of Event or Problem · 1

"DURING CORONARY ARTERY BY-PASS GRAFTING G-6000 APPLIED WITH APPLIER AND SPRING CLIP BROKE AT THE ROUND HUB WHERE APPLIER IS PLACED - HUB AND SPRING CAME APART - ALL PIECES RETRIEVED FROM PT."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SURGICAL SPRING CLIPS-DOUBLE SOFT, 6MM SURGICAL SPRING CLIPS DSS APPLIED MEDICAL RESOURCES G-6000 1024039

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other