FDA Adverse Event Malfunction Summary report: N

VASOPRESS DIGITAL MINI PUMP

MDR report key: 9582810 · Received January 13, 2020

Report

Report Number
0001526350-2020-00055
Event Type
Malfunction
Date Received
January 13, 2020
Report Date
January 13, 2020
Manufacturer
ZIMMER SURGICAL, INC.
Product Code
JOW
PMA / PMN Number
K101915
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE PRODUCT CODE: JOW. (B)(4). COULD NOT CONFIRM THE ALARM ISSUE. REPLACED THE DAMAGED POWER CORD AND FACE LABEL. THE REPORTED EVENT, "CP/OP" IS NOT CONFIRMED AND "POWER CORD DAMAGE AND OPEN WIRES" ARE CONFIRMED. THE REPORTED EVENT "CP/OP" WAS NOT CONFIRMED SINCE THE DEVICE FUNCTIONED AS INTENDED DURING PRODUCT REVIEW; HENCE, A ROOT CAUSE CANNOT BE DETERMINED. THE INVESTIGATION WAS BASED ON THE INFORMATION THAT WAS PROVIDED INITIALLY AND ANY INFORMATION THAT WAS OBTAINED THROUGHOUT THE FOLLOW-UP PROCESS. DAMAGE TO THE POWER CORD CAN OCCUR BY THE CORD/PLUG MOLD BEING RUN OVER BY WHEELED EQUIPMENT OR BY THE CORD BEING PULLED FROM THE WALL OUTLET INADVERTENTLY. IN SOME CASES, DAMAGE TO THE WIRE COVERINGS AND INSULATION MAY EXPOSE THE WIRES OF THE CORD WHEN EXCESSIVE FORCE IS APPLIED IN A CLINICAL SETTING. UNDER NORMAL CONDITIONS THE POWER CORD AND PLUG MOLD ARE NOT LIKELY TO BECOME DAMAGED. FOR VP500DM: THE AVERAGE AGE OF THE POWER CORD DURING REPLACEMENT IS 2.5 YEARS. THE FAILURE RATE FOR THE POWER CORD IS (B)(4)%. REVIEW OF THE INFORMATION PROVIDED DURING THE INVESTIGATION DETERMINED THERE IS NO FURTHER ACTIONS NEEDED AT THIS TIME (IE/CAPA/SCAR/HHE/D). THIS IS A WELL-KNOWN FAILURE MODE, WITH NO ALLEGATIONS OF HARM OR INJURY. THIS COMPLAINT WILL BE TRACKED AND TRENDED FOR ANY ADVERSE TRENDS THAT MAY WARRANT FURTHER ACTION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WERE EXPOSED/FRAYED WIRES. NO ADVERSE EVENTS WERE REPORTED AS A RESULT OF THIS MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
47551 VASOPRESS DIGITAL MINI PUMP SLEEVE, LIMB, COMPRESSIBLE JOW ZIMMER SURGICAL, INC. N/A UNK

Patients

Seq Age Sex Outcome Treatment
1