MORPHEUS CT PICC
Report
- Report Number
- 1319211-2007-00012
- Event Type
- Malfunction
- Date Received
- April 27, 2007
- Date of Event
- March 27, 2007
- Report Date
- April 27, 2007
- Manufacturer
- ANGIODYNAMICS, INC.
- Product Code
- LJS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
LOT HISTORY RECORD REVIEW: THE LOT HISTORY RECORDS WERE REVIEWED FOR ANY ABNORMALITIES THAT MAY HAVE CONTRIBUTED TO THE CAUSE OF THE COMPLAINT. NOTHING KNOWN TO HAVE CONTRIBUTED TO THE COMPLAINT WAS OBSERVED. THE GUIDEWIRE USED IN THIS MFG LOT IS A LAKE REGION GUIDEWIRE, CATALOG # 06508704, INCOMING RECEIVER # 2006137404, VENDOR LOT# P1-982868. THIS LOT PASSED ALL SPEC AT INCOMING. NOTHING KNOWN TO HAVE CONTRIBUTED TO THE COMPLAINT WAS OBSERVED. REVIEW OF RETURNED SAMPLE: ONLY THE WIRE WAS RETURNED. THE WIRE WAS IN TWO PIECES. IT APPEARS AS IF THE WIRE STRETCHED UNTIL IT BROKE. THE COMPLAINT SAMPLE AND INFO WERE FORWARDED TO THE VENDOR FOR FURTHER EVAL. CONCLUSION: THE COMPLAINT INVESTIGATION HAS BEEN CONFIRMED DURING THE VISUAL INSPECTION OF THE RETURNED SAMPLE. THE VISUAL EXAMINATION OF THE RETURNED SAMPLE SHOWED THE COIL SEGMENT SEPARATED FROM THE SHAFT DUE TO A FRACTURE OF THE CORE SHAFT IMMEDIATELY DISTAL OF THE PROXIMAL COIL TO CORE WIRE JOINT PRESENTING INDICATIONS OF BENDING OVERLOAD WITH SOME TORSIONAL LOADING. THE DISTAL ASPECT OF THE COIL WIRE IS STRETCHED. THE BEND DAMAGE APPEARS TO BE CONSISTENT WITH MULTIPLE FAILED ATTEMPTS TO ADVANCE AND/OR RETRACT THE SPECIMEN AGAINST A CONSTRAINT OR OBSTRUCTION. THE EXACT CAUSE OF THE COMPLAINT IS UNK. A REVIEW OF THE MFG, INSPECTING AND PACKAGING RECORDS INDICATED THE LOT WAS MFG AND RELEASED TO SPEC. BASED ON THE EVAL OF THE RETURNED SAMPLE, IT APPEARS AS IF PROCEDURAL AND CLINICAL FACTORS MAY HAVE IMPACTED THIS REPORTED COMPLAINT. HOWEVER, ANGIODYNAMICS HAS BEEN MADE AWARE OF THIS COMPLAINT AND HAS OPENED UP A CORRECTIVE ACTION TO ADDRESS THIS ISSUE. WE ARE CURRENTLY LOOKING INTO A NEW WIRE. THE FOLLOWING WARNINGS ARE IN THE INSTRUCTIONS FOR USE TO HELP AVOID THIS TYPE OF COMPLAINT: "NEVER USE FORCE TO REMOVE THE .018" GUIDEWIRE. RESISTANCE CAN DAMAGE THE CATHETER. IF RESISTANCE OR BUNCHING OF THE CATHETER IS OBSERVED, STOP GUIDEWIRE WITHDRAW AND ALLOW THE CATHETER TO RETURN TO NORMAL SHAPE. WITHDRAW BOTH THE CATHETER AND GUIDEWIRE TOGETHER APPROX 2CM AND REATTEMPT GUIDEWIRE REMOVAL. REPEAT THIS PROCEDURE UNTIL THE GUIDEWIRE IS EASILY REMOVED. ONCE THE GUIDEWIRE IS OUT. ADVANCE THE CATHETER INTO THE DESIRED POSITION (ZERO MARK)." "DO NOT ADVANCE THE GUIDEWIRE PAST THE AXILLA WITHOUT FLUOROSCOPIC GUIDANCE." "LEAVING THE GUIDEWIRE IN PLACE, WITHDRAW THE NEEDLE." THIS TYPE OF COMPLAINT WILL CONTINUE TO BE MONITORED FOR TRENDS. NO FURTHER ACTION IS REQUIRED AT THIS TIME.
THE WIRE BECAME STUCK AND FLOPPY END SHEARED AWAY FROM CORE OF WIRE. A NEW WIRE WAS USED AND THE PT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MORPHEUS CT PICC | PERIPHERALLY INSERTED CENTRAL CATHETER | LJS | ANGIODYNAMICS, INC. | * | 922163 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |