INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT
Report
- Report Number
- 9611451-2020-00044
- Event Type
- Malfunction
- Date Received
- January 9, 2020
- Date of Event
- December 11, 2019
- Report Date
- December 13, 2019
- Product Code
- BZE
- PMA / PMN Number
- SEE H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). THE RT268 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K034026. METHOD: SIX PRESSURE LINES OF RT268 INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT WERE RECEIVED AT FISHER & PAYKEL HEALTHCARE IN NEW ZEALAND FOR EVALUATION AND WERE VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION OF PRESSURE LINES REVEALED THAT A MOULDING DEFECT WAS FOUND ON THE ELBOW OF PRESSURE LINES. CONCLUSION: WE WERE UNABLE TO DETERMINE WHAT CAUSED THE MOULDING DEFECT OF THE PRESSURE LINE ELBOW. THE USER INSTRUCTIONS THAT ACCOMPANY THE RT265 INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT STATES: "CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." "SET APPROPRIATE VENTILATOR ALARMS."
A HEALTHCARE FACILITY IN JAPAN REPORTED VIA A FISHER & PAYKEL HEALTHCARE FIELD REPRESENTATIVE THAT THE PRESSURE LINE OF AN RT268 INFANT EVAQUA2 BREATHING CIRCUIT FALLS OFF DURING SETUP. THERE WAS NO PATIENT INVOLVEMENT.
(B)(4). THE RT268 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K034026. THE COMPLAINT RT268 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT IS CURRENTLY EN ROUTE TO FISHSER & PAYKEL HEALTHCARE IN (B)(4) FOR EVALUATION TO DETERMINE IF IT HAD A MALFUNCTION, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. WE WILL PROVIDE A FOLLOW-UP REPORT ONCE WE HAVE COMPLETED OUR INVESTIGATION.
A HEALTHCARE FACILITY IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE FIELD REPRESENTATIVE THAT THE PRESSURE LINE OF AN RT268 INFANT EVAQUA2 BREATHING CIRCUIT FALLS OFF DURING SETUP. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 35956 | INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT | BZE | BZE | RT268 | 2100856945 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |