FDA Adverse Event Malfunction Summary report: N

INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT

MDR report key: 9573279 · Received January 9, 2020

Report

Report Number
9611451-2020-00044
Event Type
Malfunction
Date Received
January 9, 2020
Date of Event
December 11, 2019
Report Date
December 13, 2019
Product Code
BZE
PMA / PMN Number
SEE H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 0

(B)(4). THE RT268 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K034026. METHOD: SIX PRESSURE LINES OF RT268 INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT WERE RECEIVED AT FISHER & PAYKEL HEALTHCARE IN NEW ZEALAND FOR EVALUATION AND WERE VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION OF PRESSURE LINES REVEALED THAT A MOULDING DEFECT WAS FOUND ON THE ELBOW OF PRESSURE LINES. CONCLUSION: WE WERE UNABLE TO DETERMINE WHAT CAUSED THE MOULDING DEFECT OF THE PRESSURE LINE ELBOW. THE USER INSTRUCTIONS THAT ACCOMPANY THE RT265 INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT STATES: "CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." "SET APPROPRIATE VENTILATOR ALARMS."

Description of Event or Problem · 0

A HEALTHCARE FACILITY IN JAPAN REPORTED VIA A FISHER & PAYKEL HEALTHCARE FIELD REPRESENTATIVE THAT THE PRESSURE LINE OF AN RT268 INFANT EVAQUA2 BREATHING CIRCUIT FALLS OFF DURING SETUP. THERE WAS NO PATIENT INVOLVEMENT.

Additional Manufacturer Narrative · 1

(B)(4). THE RT268 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K034026. THE COMPLAINT RT268 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT IS CURRENTLY EN ROUTE TO FISHSER & PAYKEL HEALTHCARE IN (B)(4) FOR EVALUATION TO DETERMINE IF IT HAD A MALFUNCTION, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. WE WILL PROVIDE A FOLLOW-UP REPORT ONCE WE HAVE COMPLETED OUR INVESTIGATION.

Description of Event or Problem · 1

A HEALTHCARE FACILITY IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE FIELD REPRESENTATIVE THAT THE PRESSURE LINE OF AN RT268 INFANT EVAQUA2 BREATHING CIRCUIT FALLS OFF DURING SETUP. THERE WAS NO PATIENT INVOLVEMENT.