FDA Adverse Event Death Summary report: N

MULTI VIEW WORKSTATION

MDR report key: 957093 · Received November 30, 2007

Report

Report Number
1220063-2007-00038
Event Type
Death
Date Received
November 30, 2007
Date of Event
November 14, 2007
Report Date
November 30, 2007
Manufacturer
DRAEGER MEDICAL SYSTEMS, INC.
Product Code
DSI
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

WE HAVE REQUESTED THE MODEL AND SERIAL NUMBERS OF THE CITED DEVICES AND THE SOFTWARE VERSIONS INVOLVED IN THE REPORTED EVENT, WHICH IS CURRENTLY UNDER INVESTIGATION.

Description of Event or Problem · 1

DRAEGER MEDICAL SYSTEMS, INC. HAS RECEIVED INFORMATION FROM A CUSTOMER THAT A PATIENT WAS SET UP WITH TELEMETRY MONITORING BY A CARDIOLOGIST TO MEASURE HEART FREQUENCY. THE PATIENT'S TELEMETRY WAS SENDING ALARMS UNINTERRUPTEDLY FOR DIFFERENT DISTURBANCES AND THE RESPONSIBLE NURSE HAD TO REPLACE/CORRECT THE ELECTRODES A NUMBER OF TIMES. THE NURSE PUT THE TELEMETRY IN STANDBY MODE ON THE CENTRAL STATION. THE NURSE GOT INTERRUPTED AND FORGOT THE PATIENT AND THE TELEMETRY BED WAS LEFT IN STANDBY MODE. THE NURSE FOUND THE PATIENT LATER WITH A VENTRICULAR FIBRILLATION. THE TELEMETRY BED IS STILL IN STANDBY MODE. THE PATIENT PASSES AWAY AFTER UNSUCCESSFUL HEART SAVING ACTIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MULTI VIEW WORKSTATION CENTRAL STATION PATIENT MONITORING SYSTEM DSI DRAEGER MEDICAL SYSTEMS, INC. NI NA

Patients

Seq Age Sex Outcome Treatment
1 NI YR Death