FDA Adverse Event Other Summary report: N

RELIANCE 444 SINGLE CHAMBER WASHER/DISINFECTOR

MDR report key: 956922 · Received December 3, 2007

Report

Report Number
9680353-2007-00008
Event Type
Other
Date Received
December 3, 2007
Date of Event
October 11, 2007
Report Date
November 9, 2007
Manufacturer
STERIS CANADA CORPORATION
Product Code
FLE
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

STAFF MEMBER REACHED INTO WASHER ON UNLOAD SIDE AS DOOR WAS CLOSING. ARM WAS CAUGHT BY DOOR AND WOULD NOT OPEN. UNLOAD DOOR OBSTRUCTION SENSING MECHANISM MALFUNCTIONING. SINGLE BOLT USED TO HOLD OBSTRUCTION SENSING MECHANISM TOO TIGHT. MECHANISM NOT ALLOWED MOVING FREELY. THE INCIDENT WAS REPORTED BY TECHNICIAN. THERE WAS NO EVIDENCE OF MISUSE OF PRODUCT BY USER AND THE UNIT WAS UNDER A MAINTENANCE CONTRACT THROUGH AGREEMENT. IT WAS LAST SERVICED IN 2007. AS CORRECTIVE ACTION, THE TECHNICIAN LOOSENED THE BOLT HOLDING OBSTRUCTION MECHANISM PULLEY. THE UNIT IS UP AND RUNNING.

Description of Event or Problem · 1

STAFF MEMBER REACHED INSIDE THE WASHER ON UNLOAD SIDE AS THE DOOR WAS CLOSING ON THE ARM. THE EXTENT OF INJURY IS A BRUISED ARM WITH SOAR AND TENDERNESS. THE INJURED PERSON RETURNED TO WORK WITHIN A DAY. WITNESS IS NOT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RELIANCE 444 SINGLE CHAMBER WASHER/DISINFECTOR WASHER/DISINFECTOR FLE STERIS CANADA CORPORATION 1401974 NA

Patients

Seq Age Sex Outcome Treatment
1 YR Other