FDA Adverse Event Injury Summary report: N

LINER: CC E CC LIGHT FLAT PE HC LINER Ø 36 / F

MDR report key: 9568608 · Received January 9, 2020

Report

Report Number
3005180920-2019-01128
Event Type
Injury
Date Received
January 9, 2020
Date of Event
December 13, 2019
Report Date
January 9, 2020
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030807688
PMA / PMN Number
K103352
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS MANAGER: DELAYED INFECTION IN CEMENTLESS THA, 8 MONTHS AFTER PRIMARY OPERATION. INFECTION IS A KNOWN POSSIBLE ADVERSE EVENT FOLLOWING EVERY SURGERY, INCLUDING THA'S. TO DATE, THERE IS NO REASON TO SUSPECT THAT THE CAUSE MAY BE LINKED TO THE IMPLANTED DEVICES.

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 7 JANUARY 2020: LOT 179315: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 11-APR-2018. EXPIRATION DATE: 2023-03-26. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. ADDITIONAL IMPLANTS INVOLVED IN THE EVENT, BATCH REVIEW PERFORMED ON 7 JANUARY 2020: BALL HEADS: MECTACER 01.29.209 BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 36 SIZE M 0 (K112115) LOT. 179000. LOT 179000: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 3-MAY-2018. EXPIRATION DATE: 2023-04-22. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. STEM: QUADRA-H 01.12.028 CEMENTLESS, HA COATED STD STEM SIZE 8 (K082792) LOT. 185959 LOT 185959: 32 ITEMS MANUFACTURED AND RELEASED ON 11-DIC-2018. EXPIRATION DATE: 2023-11-27. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. CUP: VERSAFITCUP CC TRIO 01.26.45.1160 ACETABULAR SHELL CC TRIO NO-HOLE Ø 60 (K122911) LOT. 121904A. LOT 121904A: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 9-APR-2018. EXPIRATION DATE: 2023-03-25. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

REVISION OF ALL HIP COMPONENTS, 8 MONTHS AFTER PRIMARY SURGERY, DUE TO STREPTOCOCCOSIS INFECTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
32651 LINER: CC E CC LIGHT FLAT PE HC LINER Ø 36 / F FLAT PE HC LINER LZO MEDACTA INTERNATIONAL SA 179315 07630030807688

Patients

Seq Age Sex Outcome Treatment
1 68 YR Required Intervention