FDA Adverse Event Malfunction Summary report: N

LIFEPAK 20 DEFIBRILLATOR/MONITOR

MDR report key: 956798 · Received November 29, 2007

Report

Report Number
3015876-2007-00533
Event Type
Malfunction
Date Received
November 29, 2007
Report Date
November 2, 2007
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
LDD
PMA / PMN Number
K012274
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL, INC. EVALUATED THE DEVICE. THE ROOT CAUSE WAS DETERMINED TO BE DUE TO A FAILURE OF THE POWER SUPPLY ASSEMBLY. AFTER REPLACING THE POWER SUPPLY ASSEMBLY, PROPER OPERATION WAS CONFIRMED AND THE DEVICE WAS RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE UNIT HAS NO DC OPERATION. THERE WAS NO PT ASSOCIATED WITH THE REPORTED EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK 20 DEFIBRILLATOR/MONITOR LDD PHYSIO-CONTROL, INC. 20 NA

Patients

Seq Age Sex Outcome Treatment
1 NA YR