PRISM HCV
Report
- Report Number
- 1415939-2020-00006
- Event Type
- Malfunction
- Date Received
- January 8, 2020
- Report Date
- February 24, 2020
- Manufacturer
- ABBOTT LABORATORIES
- Product Code
- QHM
- UDI-DI
- 00380740062996
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
FURTHER INVESTIGATION OF THE CUSTOMER ISSUE INCLUDED A REVIEW OF THE COMPLAINT TEXT, A SEARCH FOR SIMILAR COMPLAINTS, REVIEW OF FIELD DATA, AND A REVIEW OF LABELING. NO ADVERSE TREND WAS IDENTIFIED FOR THE CUSTOMER'S ISSUE. NO RETURN PATIENT SAMPLE WAS AVAILABLE. HISTORICAL PERFORMANCE OF THE REAGENT LOT WAS EVALUATED USING WORLD WIDE DATA. THE PATIENT DATA WAS ANALYZED AND WITHIN EXPECTED LIMITS. LABELING WAS REVIEWED AND FOUND TO BE ADEQUATE. BASED ON ALL AVAILABLE INFORMATION AND ABBOTT DIAGNOSTICS' COMPLAINT INVESTIGATION NO PRODUCT DEFICIENCY WAS IDENTIFIED.
AN ERROR WAS IDENTIFIED ON FEBRUARY 13, 2020. THE INCORRECT PRODUCT CODE OF MZO (SECTION D.2B) WAS USED IN THE PREVIOUS SUBMISSIONS. THE CORRECT CODE IS QHM, THIS REPORT IS BEING SUBMITTED TO CORRECT THE PRODUCT CODE. NO FURTHER CHANGES OR CORRECTIONS TO THE PRODUCT INFORMATION OR EVALUATION ARE NEEDED.
ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE. AN EVALUATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.
THE CUSTOMER OBSERVED (B)(6) RESULTS ON THE PRISM ANALYZER. TWENTY-SEVEN (27) RESULTS WERE REPEAT (B)(6) AND CONFIRMED (B)(6) ON NAT. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 29107 | PRISM HCV | ANTI-HCV | QHM | ABBOTT LABORATORIES | 05046M700 | 00380740062996 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | (B)(6)| (B)(6)| (B)(6)| PRISM ANALYZERS LN 06A36-04| PRISM ANALYZERS LN 06A36-04| PRISM ANALYZERS LN 06A36-04| SNS (B)(6)| SNS (B)(6)| SNS (B)(6) |