HF UNIT "ESG-400"
Report
- Report Number
- 9610773-2020-00041
- Event Type
- Injury
- Date Received
- January 8, 2020
- Date of Event
- December 24, 2019
- Report Date
- February 17, 2020
- Manufacturer
- OLYMPUS WINTER & IBE GMBH
- Product Code
- GEI
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
THE SUSPECT MEDICAL DEVICE WAS NOT RETURNED TO OLYMPUS FOR EVALUATION/INVESTIGATION. THEREFORE, THE INVESTIGATION/EVALUATION WAS PERFORMED EXCLUSIVELY ON THE BASIS OF THE INFORMATION PROVIDED BY THE CUSTOMER. THERE WERE NO REPORTS OF ANY MALFUNCTION OF THE GENERATOR. BASED ON THE INFORMATION AVAILABLE, THE EXACT CAUSE OF THE REPORTED PHENOMENON AND THE PATIENT¿S OUTCOME COULD NOT BE DETERMINED AND IS BEING JUDGED AS UNKNOWN. HOWEVER, A MATERIAL OR QUALITY PROBLEM CAN BE EXCLUDED SINCE A MANUFACTURING AND QUALITY CONTROL REVIEW WAS PERFORMED FOR THE AFFECTED SERIAL NUMBER OF THE HF GENERATOR WITHOUT SHOWING ANY ABNORMALITIES. THE CASE WILL BE CLOSED ON OLYMPUS SIDE WITH NO FURTHER ACTIONS. THE REPORTED EVENT/INCIDENT WILL BE RECORDED FOR TRENDING AND SURVEILLANCE PURPOSES. FURTHERMORE, THE USER WILL BE INFORMED ABOUT THE INVESTIGATION RESULTS.
THE SUSPECT MEDICAL DEVICE HAS NOT YET BEEN RETURNED TO OLYMPUS FOR EVALUATION/INVESTIGATION. THEREFORE, THE EXACT CAUSE OF THE PATIENT'S OUTCOME AND THE REPORTED PHENOMENON COULD NOT BE DETERMINED AND IS BEING JUDGED AS UNKNOWN. HOWEVER, IF THE SUSPECT MEDICAL DEVICE IS RETURNED FOR EVALUATION/INVESTIGATION OR ADDITIONAL SIGNIFICANT INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE UPDATED. THIS REPORT IS BEING SUBMITTED ALTHOUGH THE SUSPECT MEDICAL DEVICE IS NOT MARKETED IN THE USA. HOWEVER, A SIMILAR DEVICE IS MARKETED. MODEL # / CATALOG #: WB91051W; BRAND NAME: HF UNIT "ESG-400"; COMMON DEVICE NAME: HF-GENERATORS; 510(K): K141225; PRODUCT CODE: GEI.
OLYMPUS WAS INFORMED THAT DURING A TRANSURETHRAL RESECTION OF THE PROSTATE (TURP) PROCEDURE, THE PATIENT SUSTAINED A PERFORATION OF THE BLADDER. THE PROCEDURE WAS THEN SWITCHED TO OPEN SURGERY TO TREAT THE PERFORATION. IT WAS REPORTED THAT PATIENT HAS A THIN BLADDER MEMBRANE. THE PROCEDURE WAS COMPLETED WITH THE SAME SET OF EQUIPMENT AND THERE WAS NO ALLEGED MALFUNCTION OF THE OLYMPUS MEDICAL DEVICES USED DURING THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 28711 | HF UNIT "ESG-400" | HF-GENERATORS | GEI | OLYMPUS WINTER & IBE GMBH | WB91051J |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |