ECHOTIP PROCORE ENDOBRONCHIAL HD BIOPSY NEEDLE

MDR report key: 9563196 · Received January 8, 2020

Report

Report Number
3001845648-2020-00014
Event Type
Malfunction
Date Received
January 8, 2020
Date of Event
December 9, 2019
Report Date
December 10, 2019
Manufacturer
COOK IRELAND LTD
Product Code
FCG
UDI-DI
00827002342798
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION IS STILL PENDING. A FOLLOW UP MDR WILL BE SUBMITTED TO INCLUDE THE INVESTIGATION CONCLUSIONS. PMA/510(K) #: K160229.

Additional Manufacturer Narrative · 0

PMA/510(K) #: K160229. 1 UNIT OF LOT C1647316 OF ECHO-HD-22-EBUS-P-C WAS RETURNED OPENED IN ITS ORIGINAL PACKAGING. THE DEVICE INVOLVED IN THE COMPLAINT WAS EVALUATED IN THE LABORATORY. THE NEEDLE WAS FOUND TO BE BROKEN DISTALLY. DAMAGE TO THE DISTAL PART OF THE SHEATH WAS ALSO OBSERVED. BASED ON WORST CASE IT IS ASSUMED THAT THE NEEDLE BROKE INSIDE THE PATIENT AND WAS REMOVED, HOWEVER, WE DON¿T NOT HAVE CONFIRMATION OF THIS INFORMATION, WE DO KNOW ¿THE MISSING PART HAS BEEN DISPOSED OF BY THE CUSTOMER¿, WE ALSO KNOW THAT NO ADDITIONAL PROCEDURE WAS REQUIRED FROM THE INFORMATION PROVIDED. PRIOR TO DISTRIBUTION, ALL ECHO-HD-22-EBUS-P-C DEVICES ARE SUBJECTED TO FUNCTIONAL CHECKS AND VISUAL INSPECTION TO ENSURE DEVICE INTEGRITY. THESE INSPECTIONS AND FUNCTIONAL CHECKS ARE OUTLINED IN INTERNAL PROCEDURES IN PLACE AT CIRL. A REVIEW OF THE MANUFACTURING RECORDS FOR ECHO-HD-22-EBUS-P-C OF LOT NUMBER C1647316 DID NOT REVEAL ANY DISCREPANCIES THAT COULD HAVE CONTRIBUTED TO THIS COMPLAINT ISSUE. THE REVIEW OF RELEVANT MANUFACTURING RECORDS, CONFIRMS THE FAILURE MODE HAS NOT PREVIOUSLY OCCURRED WITH THE CURRENT LOT NUMBER. BASED ON THE INFORMATION AVAILABLE TO DATE, THERE IS NO EVIDENCE TO SUGGEST THAT THERE ARE ANY MANUFACTURING ISSUES ASSOCIATED WITH LOT NUMBER C1647316. THE NOTES SECTION OF THE INSTRUCTIONS FOR USE, WHICH ACCOMPANIES THIS DEVICE INSTRUCTS THE USER TO; "VISUALLY INSPECT WITH PARTICULAR ATTENTION TO KINKS, BENDS AND BREAKS. IF AN ABNORMALITY IS DETECTED THAT WOULD PROHIBIT PROPER WORKING CONDITION, DO NOT USE" AND "ENSURE THE STYLET IS FULLY INSERTED WHEN ADVANCING THE NEEDLE INTO THE BIOPSY SITE". THERE IS EVIDENCE TO SUGGEST THAT THE CUSTOMER DID NOT FOLLOW THE INSTRUCTIONS FOR USE IN RELATION TO THE USE OF THE STYLET. A DEFINITIVE ROOT CAUSE FOR THE CUSTOMER COMPLAINT COULD NOT BE DETERMINED AS CIRCUMSTANCES OF USE CANNOT BE REPLICATED IN THE LABORATORY. A POSSIBLE ROOT CAUSE COULD BE ATTRIBUTED TO USER ERROR AS THE STYLET SHOULD NOT BE PARTIALLY REMOVED PRIOR TO ADVANCEMENT OF THE NEEDLE INTO INTENDED TARGETED SITE. THIS MAY HAVE CAUSED THE NEEDLE BREAKAGE AND POSSIBLY THEN RESULTED IN THE DISTAL SHEATH DAMAGE. COMPLAINT IS CONFIRMED AS THE FAILURE WAS VERIFIED IN THE LABORATORY. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE. THE MISSING PART OF THE NEEDLE WAS DISPOSED OF BY THE CUSTOMER. COMPLAINTS OF THIS NATURE WILL CONTINUE TO BE MONITORED FOR POTENTIAL EMERGING TRENDS.

Description of Event or Problem · 0

NEEDLE WENT TROUGH THE DISTAL PART SHEATH. AS PER CC FORM: WHEN THE NEEDLE WAS PUSHED FORWARD, THE CATHETER WAS PIERCED. THAT WAS NOT ANGLED AT THE TIME. THE ENDOSCOPE AND THE PATIENT REMAINED UNHARMED. THE COMPLAINT DEVICE WAS RETURNED AND EVALUATED ON (B)(6) 2020 AND THE DISTAL END OF THE NEEDLE WAS CONFIRMED TO BE BROKEN. NO PART OF THE NEEDLE REMAINS IN THE PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED TO CONFIRM IF THE NEEDLE BROKE INSIDE / OUTSIDE THE PATIENT.

Description of Event or Problem · 0

NEEDLE WENT TROUGH THE DISTAL PART SHEATH. AS PER CC FORM: WHEN THE NEEDLE WAS PUSHED FORWARD, THE CATHETER WAS PIERCED. THAT WAS NOT ANGLED AT THE TIME. THE ENDOSCOPE AND THE PATIENT REMAINED UNHARMED.

Additional Manufacturer Narrative · 1

INVESTIGATION IS STILL PENDING. A FOLLOW UP MDR WILL BE SUBMITTED TO INCLUDE THE INVESTIGATION CONCLUSIONS. PMA/510(K) #: K160229.

Description of Event or Problem · 1

NEEDLE WENT TROUGH THE DISTAL PART SHEATH. AS PER CC FORM: WHEN THE NEEDLE WAS PUSHED FORWARD, THE CATHETER WAS PIERCED. THAT WAS NOT ANGLED AT THE TIME. THE ENDOSCOPE AND THE PATIENT REMAINED UNHARMED.