FDA Adverse Event Malfunction Summary report: N

TRUE METRIX

MDR report key: 9560417 · Received January 7, 2020

Report

Report Number
1000113657-2020-00010
Event Type
Malfunction
Date Received
January 7, 2020
Date of Event
December 9, 2019
Report Date
January 7, 2020
Manufacturer
TRIVIDIA HEALTH INC
Product Code
NBW
UDI-DI
00021292007836
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INTERNAL REPORT REFERENCE NUMBER: (B)(4). METER WAS RETURNED FOR EVALUATION. NO DEFECT WAS DETECTED. TEST STRIPS WERE NOT RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-20: USER'S TEST STRIP HAD POOR STORAGE. NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - UNABLE TO ESTABLISH CONTACT WITH CUSTOMER AT THIS TIME. PRODUCT NOTIFICATION LETTER SENT TO CONTACT CUSTOMER CARE.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS OBTAINED OF 233, 171, 218, 226 AND 281 MG/DL. THE CUSTOMER¿S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT IS 130 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION AND IS STORED IN THE KITCHEN. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 06/30/2021. CUSTOMER DID NOT HAVE ANOTHER VIAL OF TEST STRIPS THAT HAD BEEN STORED AND HANDLED CORRECTLY. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: RESULT 1: 233 MG/DL DATE: (B)(6) 2019 TIME: 8:05 AM FASTING . RESULT 2: 171 MG/DL DATE: (B)(6) 2019 TIME: 4:48 PM FASTING . RESULT 3: 218 MG/DL DATE: (B)(6) 2019 TIME: 3:58 PM FASTING . RESULT 4: 226 MG/DL DATE: (B)(6) 2019 TIME: 3:42 PM FASTING . RESULT 5: 281 MG/DL DATE: (B)(6) 2019 TIME: 3:20 PM FASTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
22521 TRUE METRIX SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER NBW TRIVIDIA HEALTH INC STRIP, TMX WGN 100CT12/CS MG/DL #383569 MW4028S 00021292007836

Patients

Seq Age Sex Outcome Treatment
1