BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM
Report
- Report Number
- 3006948883-2020-00001
- Event Type
- Malfunction
- Date Received
- January 7, 2020
- Date of Event
- November 19, 2019
- Report Date
- January 17, 2020
- Manufacturer
- BD (SUZHOU)
- Product Code
- FOZ
- PMA / PMN Number
- PENDING
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.6. INVESTIGATION: A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 9141617. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. UNFORTUNATELY A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING. WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT.
IT WAS REPORTED THAT PRIOR TO USE IT WAS DISCOVERED THAT THE NEEDLE WAS LOOSE WITH A BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: BEFORE PENETRATION AFTER UNWRAPPED THE PACKAGE IT'S NOTICED THAT NEEDLE CORE AND SAFETY SHIELD LOOSEN. THUS UNABLE TO PENETRATE.
(B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT PRIOR TO USE IT WAS DISCOVERED THAT THE NEEDLE WAS LOOSE WITH A BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: BEFORE PENETRATION AFTER UNWRAPPED THE PACKAGE IT'S NOTICED THAT NEEDLE CORE AND SAFETY SHIELD LOOSEN. THUS UNABLE TO PENETRATE.