SILKAM BLACK 0 (3.5) 75CM HS30
Report
- Report Number
- 3003639970-2019-00960
- Event Type
- Malfunction
- Date Received
- January 7, 2020
- Date of Event
- December 17, 2019
- Report Date
- March 2, 2020
- Manufacturer
- B.BRAUN SURGICAL SA
- Product Code
- GAP
- PMA / PMN Number
- SEE H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- RS
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION: SAMPLES RECEIVED: 30 UNOPENED POUCHES. ANALYSIS AND RESULTS: THERE ARE NO PREVIOUS COMPLAINTS OF THIS CODE-BATCH. WE MANUFACTURED AND DISTRIBUTED IN THE MARKET (B)(4) UNITS OF THIS CODE-BATCH. THERE ARE NO UNITS IN OUR STOCK. WE HAVE RECEIVED 30 CLOSED SAMPLES FOR ANALYSIS. WE HAVE TESTED THE KNOT PULL TENSILE STRENGTH OF ALL SAMPLES RECEIVED AND THE RESULTS FULFIL THE REQUIREMENTS OF THE EUROPEAN PHARMACOPOEIA (EP): 2.60 KGF IN AVERAGE AND 2.45 KGF IN MINIMUM (EP REQUIREMENTS: 2.24 KGF IN AVERAGE AND 1.33 KGF IN MINIMUM). REVIEWED THE BATCH MANUFACTURING RECORD, THIS PRODUCT HAD A NORMAL PROCESS AND THE RESULTS DURING THE PROCESS FULFIL USP/EP AND B.BRAUN SURGICAL REQUIREMENTS. REMARKS: WHEN WORKING WITH SILKAM GREAT CARE MUST BE TAKEN TO ENSURE THAT THE USE OF SURGICAL INSTRUMENTS, SUCH AS TWEEZERS OR NEEDLE HOLDER, DOES NOT LEAD TO CRIMPING DAMAGE OF THE SUTURE. FINAL CONCLUSION: ALTHOUGH THE RESULTS OF THE SAMPLES RECEIVED FULFIL THE SPECIFICATIONS OF EUROPEAN PHARMACOPOEIA/ B. BRAUN SURGICAL SPECIFICATIONS, WE TAKE NOTE OF THIS INCIDENCE IN ORDER TO ASSESS IF NEW OR ADDITIONAL ACTIONS ARE NEEDED. ACTIONS ON PRODUCT DISTRIBUTED OF THIS REFERENCE/BATCH: BASED ON THE CONCLUSION DERIVED FROM INVESTIGATION, IT IS NOT REQUIRED TO MAKE ACTIONS IN DISTRIBUTED PRODUCT. NO CORRECTIVE/PREVENTIVE ACTIONS NEEDED.
PMA/510K: REPORTED DEVICE NOT MARKETED IN THE U.S., HOWEVER, SIMILAR DEVICES OR DEVICES THAT SHARE COMPONENTS, RAW MATERIALS, PROCESS METHODS OR OTHER TECHNOLOGICAL CHARACTERISTICS ARE REGISTERED WITHIN THE U.S. K990089. IF ADDITIONAL INFORMATION BECOMES AVAILABLE A FOLLOW UP REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT THE THREADS BREAK. THE REPORTER INDICATED THAT THE THREAD BROKE DURING SURGERY. PATIENT INFORMATION IS NOT AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 20518 | SILKAM BLACK 0 (3.5) 75CM HS30 | OTHER SUTURE | GAP | B.BRAUN SURGICAL SA | C0764620 | 619224 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |