SJM MASTERS SERIES HEMODYNAMIC PLUS VALVE
Report
- Report Number
- 3007113487-2019-00039
- Event Type
- Injury
- Date Received
- January 7, 2020
- Date of Event
- December 11, 2019
- Report Date
- February 18, 2020
- Manufacturer
- ST. JUDE MEDICAL CATD
- Product Code
- LWQ
- UDI-DI
- 05415067016522
- PMA / PMN Number
- P810002
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- PHYSICIAN
Narratives
AN EVENT OF LEAFLET IMMOBILITY AND INCREASE OF GRADIENT DUE TO POSSIBLE THROMBUS IN A PATIENT WITH MULTIPLE CONCOMITANT INFECTIONS DUE TO COMORBIDITIES AND SUSPECTED ENDOCARDITIS WAS REPORTED. THE RESULTS OF THE INVESTIGATION ARE INCONCLUSIVE SINCE THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE DEVICE HISTORY RECORD WAS REVIEWED TO ENSURE THAT EACH MANUFACTURING AND INSPECTION OPERATION WAS PERFORMED AND THE PRODUCT MET ALL SPECIFICATIONS AT THE TIME OF COMMERCIALIZATION. BASED ON THE INFORMATION RECEIVED, THE CAUSE OF THE REPORTED INCIDENT COULD NOT BE CONCLUSIVELY DETERMINED.
AFTER MHV IMPLANTATION ON (B)(6) 2019 OF A 15MHPJ-505 (SN (B)(6) ) UNDER COMPASSIONATE USE (BEFORE CE MARK), IT WAS DETECTED ON (B)(6) 2019 A BLOCK IN ONE OF THE LEAFLETS (NOT MOVING), AFTER DETECTION OF LOW PATIENT OUTPUT AND INCREASE OF GRADIENTS. ANTITROMBOTIC THERAPY WAS INTRODUCED WITHOUT SUCCESS FACING A POSSIBLE THROMBUS. THE PATIENT WASN¿T UNDER AC SINCE INTERVENTION BECAUSE OF VERY YOUNG AGE AND MULTIPLE COMORBIDITIES. THE PATIENT HAD BEEN UNDER LOW MOLECULAR WEIGHT HEPARYN. THE PATIENT HAS HAD LOT OF CONCOMITANT INFECTIONS DUE TO COMORBIDITIES. SURGEONS SUSPECT ON ENDOCARDITIS. THE PROSTHESIS HAS BEEN EXPLANTED TODAY (B)(6) ) AND SENT BY THE SURGEONS TO THE MICROBIOLOGY LAB AT THEIR HOSPITAL FOR ANALYSIS. THEY HAVE IMPLANTED ANOTHER ABBOTT 15MMS.
THE RESULTS/METHOD AND CONCLUSION CODES ALONG WITH INVESTIGATION RESULTS WILL BE PROVIDED IN A SUBSEQUENT SUBMISSION.
AFTER MHV IMPLANTATION ON (B)(6) 2019 OF A 15MHPJ-505 (SN (B)(4)) UNDER COMPASSIONATE USE (BEFORE CE MARK), IT WAS DETECTED ON (B)(6) 2019 A BLOCK IN ONE OF THE LEAFLETS (NOT MOVING), AFTER DETECTION OF LOW PATIENT OUTPUT AND INCREASE OF GRADIENTS. ANTITHROMBOTIC THERAPY WAS INTRODUCED WITHOUT SUCCESS FACING A POSSIBLE THROMBUS. THE PATIENT WASN'T UNDER AC SINCE INTERVENTION BECAUSE OF VERY YOUNG AGE AND MULTIPLE COMORBIDITIES. THE PATIENT HAD BEEN UNDER LOW MOLECULAR WEIGHT HEPARYN. THE PATIENT HAS HAD LOT OF CONCOMITANT INFECTIONS DUE TO COMORBIDITIES. SURGEONS SUSPECT ON ENDOCARDITIS. THE PROSTHESIS HAS BEEN EXPLANTED TODAY ((B)(6)) AND SENT BY THE SURGEONS TO THE MICROBIOLOGY LAB AT THEIR HOSPITAL FOR ANALYSIS. THEY HAVE IMPLANTED ANOTHER ABBOTT 15MMS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 24258 | SJM MASTERS SERIES HEMODYNAMIC PLUS VALVE | HEART-VALVE, MECHANICAL | LWQ | ST. JUDE MEDICAL CATD | 15MHPJ-505 | 5328893 | 05415067016522 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |