FDA Adverse Event Malfunction Summary report: N

PUMP MMT-1780KPK 670G PATHWAY BLACK MG

MDR report key: 9556948 · Received January 7, 2020

Report

Report Number
3004209178-2020-52094
Event Type
Malfunction
Date Received
January 7, 2020
Date of Event
December 29, 2019
Report Date
May 24, 2022
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
UDI-DI
00643169939219
Removal / Correction Number
Z-0956-2020
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS PART OF A RETROSPECTIVE REVIEW AND REMEDIATION EFFORTS IN RESPONSE TO A WARNING LETTER. UPDATED H9: Z-0956-2020. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Additional Manufacturer Narrative · 1

DEVICE PASSED THE REWIND TEST, PRIME OR SEATING TEST, BASIC OCCLUSION TEST, FORCE SENSOR TEST AND SELF TEST. DEVICE RECEIVED WITH MISSING RETAINER. DEVICE RECEIVED WITH PARTIALLY BROKEN RESERVOIR TUBE LIP. UNABLE TO PERFORM THE DISPLACEMENT TEST AND P-CAP OR RESERVOIR WILL NOT LOCK PROPERLY DUE TO MISSING RETAINER.

Description of Event or Problem · 1

CUSTOMER REPORTED VIA PHONE CALL THAT THE CLEAR RING DETACHED FROM THE INSULIN PUMP. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 256 MG/DL. CUSTOMER STATED THAT THE PLASTIC RING WHERE THE RESERVOIR GOES IT APPEARS IT FELL OFF. CUSTOMER STATED THAT THEY GLUED THE CLEAR PLASTIC RING TO BACK INSULIN PUMP. CUSTOMER STATED THAT TREATED WITH BOLUS FOR BLOOD GLUCOSE VALUE. CUSTOMER STATED AUTO MODE WAS ACTIVE AND THEY DECLINED TO TROUBLESHOOT FOR THE BLOOD GLUCOSE VALUE. CUSTOMER STATED THE INSULIN PUMP HAD COSMETIC DAMAGE. CUSTOMER STATED THE INFUSION SET AND RESERVOIR DID NOT SHOW SIGNS OF DAMAGE. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK-UP PLAN. THE DEVICE WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
23518 PUMP MMT-1780KPK 670G PATHWAY BLACK MG ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1780KPK HG1YP76 00643169939219

Patients

Seq Age Sex Outcome Treatment
1 55 YR Male