FDA Adverse Event
Malfunction
Summary report: N
VENTAK PRX AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
MDR report key: 95541
·
Received May 8, 1997
Report
- Report Number
- 2124215-1997-01244
- Event Type
- Malfunction
- Date Received
- May 8, 1997
- Date of Event
- February 12, 1997
- Report Date
- February 12, 1997
- Manufacturer
- CARDIAC PACEMAKERS
- Product Code
- LWS
- Removal / Correction Number
- N030/031-5
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- PHYSICIAN
Narratives
Description of Event or Problem · 1
EVENT DESCRIPTION CPI RECEIVED INFORMATION THAT AT A VENTAK PRX IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS FOUND TO BE IN MONITOR ONLY MODE (THERAPY DELIVERY DISABLED) WHEN INTERROGATED AT A NORMAL FOLLOW-UP VISIT. AFTER INTERVIEWING THE PATIENT, IT IS BELIEVED THAT THE ICD SWITCHED TO MONITOR ONLY MODE FROM EXPOSURE TO A MAGNETIC FIELD. NOTE: THE VENTAK PRX PHYSICIAN'S MANUAL (CPI PUBLICATION 352927) P 74 STATES: '...PATIENTS SHOULD AVOID EQUIPMENT OR SITUATIONS WHERE THEIR DEVICE WOULD BE EXPOSED TO STRONG (>10 GAUSS) MAGNETIC FIELDS.' THIS ICD FALLS UNDER FDA SAFETY ALERT FDA # N030/031-5. THE PHYSICIAN WAS NOTIFIED ON 4/18/95.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VENTAK PRX AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Implant | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | CARDIAC PACEMAKERS | 1705 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR | Other | THE DEVICE 0041/761393 WAS IMPLANTED 14-AUG-1987| THE DEVICE 0030/072207 WAS IMPLANTED 14-AUG-1987| THE DEVICE 0030/016260 WAS IMPLANTED 14-AUG-1987| THE DEVICE 0041/761394 WAS IMPLANTED 14-AUG-1987 |