FDA Adverse Event Malfunction Summary report: N

BD VENFLON¿ PRO SAFETY SHIELDED IV CATHETER

MDR report key: 9553733 · Received January 6, 2020

Report

Report Number
8041187-2019-01073
Event Type
Malfunction
Date Received
January 6, 2020
Date of Event
December 19, 2019
Report Date
January 14, 2020
Manufacturer
BECTON DICKINSON MEDICAL (SINGAPORE)
Product Code
FOZ
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: ONE VIDEO WAS RECEIVED BY OUR QUALITY TEAM FOR EVALUATION. UPON VISUAL INSPECTION OF THE VIDEO IT WAS OBSERVED THAT THE VALVE UNDER THE PORT HAD MOVED; THEREFORE THE INCIDENT COULD BE VERIFIED DUE TO LEAKAGE BEING CAUSED BY THE MOVED VALVE. A DEVICE HISTORY RECORD REVIEW FOUND NO NON-CONFORMANCES ASSOCIATED WITH THIS ISSUE DURING PRODUCTION OF THIS BATCH. BASED ON THE QUALITY TEAM'S INVESTIGATION, THE ROOT CAUSE OF THE LEAKAGE IS DUE TO THE INJECTION VALVE MOVING WITHIN THE CANNULA HUB. A TREND FOR THE MOVING INJECTION VALVE HAS BEEN OBSERVED FOR THIS PRODUCT LINE. A PROJECT HAS BEEN STARTED, CAPA#1298780, TO FURTHER DETERMINE THE REASON FOR THE MOVING INJECTION VALVE IN THE VENFLON PRO SAFETY SO THAT A FIX CAN BE MADE TO FURTHER PREVENT THIS ISSUE. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD VENFLON¿ PRO SAFETY SHIELDED IV CATHETER LEAKED BLOOD OUT OF THE OPEN INJECTION PORT AFTER THE CATHETER WAS PLACED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM GERMAN TO ENGLISH: BLOODS FLOWS OUT OF THE OPEN INJECTION PORT AFTER THE CATHETER WAS PLACED.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BD VENFLON¿ PRO SAFETY SHIELDED IV CATHETER LEAKED BLOOD OUT OF THE OPEN INJECTION PORT AFTER THE CATHETER WAS PLACED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: BLOODS FLOWS OUT OF THE OPEN INJECTION PORT AFTER THE CATHETER WAS PLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
19008 BD VENFLON¿ PRO SAFETY SHIELDED IV CATHETER INTRAVASCULAR CATHETER FOZ BECTON DICKINSON MEDICAL (SINGAPORE) 9241386

Patients

Seq Age Sex Outcome Treatment
1 Other