FDA Adverse Event Injury Summary report: N

INFINION CX

MDR report key: 9549280 · Received January 3, 2020

Report

Report Number
3006630150-2019-07811
Event Type
Injury
Date Received
January 3, 2020
Date of Event
October 21, 2019
Report Date
January 3, 2020
Manufacturer
BOSTON SCIENTIFIC NEUROMODULATION
Product Code
LGW
UDI-DI
08714729861638
PMA / PMN Number
P030017
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
ID, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

MODEL NUMBER/CATALOG NUMBER: SC-2317-70, SERIAL NUMBER: (B)(4), BATCH/LOT NUMBER: 5073230, MODEL/CATALOG DESCRIPTION: INFINION CX LEAD KIT, 70CM. THE EXPLANTED DEVICES WERE NOT RETURNED TO BSN AS THEY WERE DISCARDED BY THE MEDICAL FACILITY. IT IS INDICATED THAT THE DEVICES WILL NOT BE RETURNED FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICES COULD NOT BE COMPLETED. A REVIEW OF THE DEVICE HISTORY RECORDS WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MED WATCH WILL BE FILED.

Description of Event or Problem · 1

A REPORT WAS RECEIVED THAT THE PATIENT WAS EXPERIENCING INADEQUATE STIMULATION DUE TO LEAD MIGRATION. THE PATIENT UNDERWENT A LEAD REPLACEMENT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
11913 INFINION CX STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF LGW BOSTON SCIENTIFIC NEUROMODULATION SC-2317-70 5073172 08714729861638

Patients

Seq Age Sex Outcome Treatment
1 51 YR Required Intervention