BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM
Report
- Report Number
- 3006948883-2019-01170
- Event Type
- Malfunction
- Date Received
- January 3, 2020
- Date of Event
- November 20, 2019
- Report Date
- January 16, 2020
- Manufacturer
- BD (SUZHOU)
- Product Code
- FOZ
- PMA / PMN Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.6. INVESTIGATION SUMMARY BD HAD NOT RECEIVED SAMPLES OR PHOTOS FROM THE CUSTOMER FACILITY FOR EVALUATION; THEREFORE, THE INVESTIGATION WAS LIMITED. RETENTION SAMPLES WERE SELECTED FROM BD INVENTORY FOR EVALUATION/TESTING AND UPON COMPLETION, NO ISSUES WERE OBSERVED RELATING TO LEAKAGE AS ALL RETENTION SAMPLES MET SPECIFICATIONS. A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 9141672. OUR RECORDS SHOW THAT THIS IS THE SECOND INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. UNFORTUNATELY A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING. WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT.
IT WAS REPORTED THAT LEAKAGE OCCURRED DURING USE WITH A BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "DURING INTRAVENOUS INFUSION, THE INDWELLING NEEDLE WAS LEAKED. AFTER REMOVAL, THE NURSE WAS EXPLAINED TO THE PATIENT AND THE INDWELLING NEEDLE WAS REPLACED."
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT LEAKAGE OCCURRED DURING USE WITH A BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "DURING INTRAVENOUS INFUSION, THE INDWELLING NEEDLE WAS LEAKED. AFTER REMOVAL, THE NURSE WAS EXPLAINED TO THE PATIENT AND THE INDWELLING NEEDLE WAS REPLACED."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 12296 | BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM | INTRAVASCULAR CATHETER | FOZ | BD (SUZHOU) | 9141672 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |