FDA Adverse Event Death Summary report: N

SIGMA SPECTRUM

MDR report key: 953521 · Received July 11, 2006

Report

Report Number
1314492-2006-00004
Event Type
Death
Date Received
July 11, 2006
Date of Event
June 4, 2006
Report Date
June 12, 2006
Manufacturer
SIGMA INT'L
Product Code
FRN
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

PRODUCT WAS RECEIVED AND EXAMINED AT SIGMA 7/5/06. THE PLUG IN POWER SUPPLY DISPLAYED NO GREEN LED WHEN PLUGGED IN INDICATING NO POWER OUT. BROKEN TRACES CONSISTENT WITH TRAUMA TO THE ASSEMBLY AS A RESULT OF DROPPING WERE IDENTIFIED. WITH NO INDICATION OF POWER (GREEN LED LIT) THIS PUMP SHOULD NOT HAVE BEEN IN SERVICE.

Description of Event or Problem · 1

SIGMA SPECTRUM PUMP RETRIEVED FROM HOLDING AREA TO REPLACE PUMP ON PT WITH BATTERY DEPLETION. THIS PUMP WAS PLUGGED INTO THE WALL OUTLET IN THE HOLDING AREA. IT WOULD NOT TURN ON EVEN WHEN PLUGGED IN. CLINICAL ENGINEERS, RM NURSING EVALUATED THE PUMP 6/5/06. THIS WAS REPEATED WITH CEO OF SIGMA AND LOCAL SIGMA REP ON 6/6/06. THE PUMP WOULD NOT TURN ON. UNABLE TO DETERMINE BATTERY CAPACITY. THE STRAIN RELIEF BRACKET INTACT. ONLY ONE SCREW. THE GREEN LED LIGHT ON THE TRANSFORMER (PLUG) IS NOT ON. UNABLE TO GET A READING FROM THE BATTERY CAPACITY. ALL FINDINGS REPORTED TO SIGMA ON 6/6/06.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SIGMA SPECTRUM INFUSION PUMP FRN SIGMA INT'L SPECTRUM NA

Patients

Seq Age Sex Outcome Treatment
1 57 YR Death